WIKA operating instructions, model CTR3000
27
14191761.01 05/2019 EN/DE
6.4.1.6 Probe selection
By pressing button
13
the menu for the probe selection opens on the right side. Back via the menu button on the bottom line.
1. Press button
13
.
⇒
The menu for the probe selection opens on the right side.
2. Select the probe.
⇒
The desired probe will be indicated in the button
13
.
3. Return to the main screen by pressing the menu button
24
.
Standard probes, default
Probes for all PRT channels
3w PRT (Pt100)
Pt100, 3w, DIN conversion, internal 100 Ω, keep warm current off
3W-PT100
4w PRT (Pt100)
Pt100, 4w, DIN conversion, internal 100 Ω, keep warm current off
4W-PT100
SPRT (Pt25)
Pt25, 4w, DIN conversion, intern 25 Ω
4W-PT25
Thermistor
500 kΩ, no temperature conversion
THERMISTOR
Probes for all TC channels
Thermocouple
TC K, internal reference cold junction
Stored probes
■
List of all configured probes in the menu
“Probes”
.
■
With
[+]
allows the user to configure a new probe and the user is directly linked to the menu
“Probes”
(details in chapter
6.4.3 “Application [Probes]”).
■
Sequence of the listed probes acc. ABC.
A probe can only be assigned to one channel at a time, so it is not possible to e.g. assign stored probe
PRT1
to channel 1 and 2 at the same time. This means that if a probe, which is currently assigned to a channel, is
assigned to another channel, it is no longer assigned to the first channel. If no probe is explicitly assigned to a
channel, the default probe for the channel type will be used. For PRT channels, this is 4W-PT100, for TC channels
TC(K).
13
24
6. Commissioning, operation