Error messages

In the event of a fault or error, the unit displays an error
code in the status display.
These error codes are described in detail below 
(see also the inside of the front plate of the unit).

001:

No parts contacted/no gold electrolyte present.
Rectify the fault, then press the Start/Stop button
to activate the process.

002:

Temperature sensor defective/not correctly
contacted or heater defective.
First make sure that the temperature sensor is
inserted correctly. If this does not rectify the
fault, the temperature sensor or the heater is
defective. In this case, contact WIELAND
 TECHNIK or your authorized dealer.

003:

Maximum capacity exceeded.
If the maximum capacity (550 ml) is exceeded
when entering the power level, a further increase
of the power level is not possible. The corre-
sponding channel can be reset to zero by pres-
sing any power level button. Then remove the
excess parts.

004:

Contact fault.
One or more parts were not contacted correctly.
In addition, the power level display alternates
between the set value and zero. Rectify the fault
by pressing the Start/Stop button. Then press the
Start/Stop button once again to activate the pro-
cess. If the contact error is not rectified, the unit
will still commence the heating-up phase after
30 sec.

005:

Electronic fault.

006:

Software fault.

007:

Software fault.
In the event of an electronic fault or a fault 
in the software, please contact WIELAND
 TECHNIK or your authorized dealer.

6

6..  P

Po

os

ss

siib

blle

e  ffa

au

ulltts

s  a

an

nd

d  ttrro

ou

ub

blle

es

sh

ho

oo

ottiin

ng

g

20

Содержание AGC Micro

Страница 1: ...ellung von Kronen Ankerkronen f r kleine Br cken mit Keramik bzw Kunststoffverblendungen sowie f r Mesostrukturen SupraCaps in der Implantatprothetik Prothesenbasen und Sekund rkronen in der Doppelkro...

Страница 2: ...1 9 10 11 12 13 17 18 10 20 2 3 4 5 6 7 8 14 15 16...

Страница 3: ...it ten permanent wissenschaftlich begleitet Eine Vielzahl von unabh ngigen Ver ffentlichungen belegen die sehr gute klinische Prognose von Restaurationen die mit diesem Verfahren hergestellt wurden WI...

Страница 4: ...e des Ger tes 6 2 Kronen galvanisieren 8 3 Prothesenbasis galvanisieren 15 4 Restgoldabscheidung 17 5 Recycling 19 6 Fehlerquellen 20 7 Reinigungshinweise 24 8 Technische Daten 26 9 Abscheidungsprotok...

Страница 5: ...n das Innere des Ger tes gelangen l Kontrollieren Sie das Ger t vor dem Gebrauch auf Besch digungen Besch digte Teile m ssen vor der Benutzung von WIELAND DENTAL TECHNIK instandgesetzt oder ausgewechs...

Страница 6: ...0 Betriebs LED Tasten und Anzeigen Stromstufentaste Taste zum Einstellen der ben tigten Stromstufe f r jeden Kanal einzeln einstellbar LED Anzeige zeigt f r jeden Kontakt die eingestellte Stromstufe a...

Страница 7: ...n und ziehen Sie diesen von der Halterung ab Die mitgelieferte Anode mit Temperaturf hler bis zum Anschlag ein schieben Mit dem AGC Micro k nnen bis zu sechs Galvanoteile pro Galvanisiervorgang herges...

Страница 8: ...zur Pr parationsgrenze so gering wie m glich halten ca 1 mm Den AGC Kupferstab mit wenig AGC Sekundenkleber einkleben Kleber berschuss unbedingt entfernen Isolier schicht Bei Verwendung eines anderen...

Страница 9: ...romstufe 3 Schichtdicke 0 2 mm 49 ml 28 ml 77 ml Teil auf Stromstufe 4 Schichtdicke 0 2 mm 37 ml Einsatz 50 ml siehe angegebene F llmengen Stromstufe Schichtdicke und Goldbadmenge festlegen Die Gr e d...

Страница 10: ...Kronen k nnen diese mit meh reren AGC Kupferst ben kontaktiert werden Die Str me addieren sich dann Rechenbeispiel 3er Verblockung z B 2 Frontz hne 1 Pr molar Schichtst rke 0 2 mm 86 ml Stufe 7 37 ml...

Страница 11: ...alvanisierung und eine h chstm gliche Ausnutzung des Goldgehaltes gew hr leistet Beachten Sie das Haltbarkeitsdatum der Goldb der und die Hinweise zur Lagerung Vorsicht das Goldbad verursacht Flecken...

Страница 12: ...nd eine weitere Erh hung der Stromstufe ist nicht mehr m glich Durch Dr cken einer beliebigen Stromstufentaste wird der entsprechen de Kanal auf Null zur ckgesetzt Entfernen Sie an schlie end die berz...

Страница 13: ...nkt in der Status Anzeige die Fehlermeldung 004 Durch Dr cken der Start Stop Taste k nnen Sie den Prozess unterbrechen und den Kontaktierungsfehler beheben z B AGC Kupferstab nicht richtig in Buchse e...

Страница 14: ...alten Mit der Start Stop Taste wird der Fehlerhinweis gel scht Anschlie end muss das Ger t ber die Timer Taste zur ckgesetzt werden Falls die Schichtdicke der Teile unter 0 15 mm liegt raten wir von e...

Страница 15: ...eben und Kupferst be oben auf der Halterung kontaktieren siehe Skizze Darauf achten dass die Prothesenbasis nicht an der Anode anliegt Timer Taste dr cken f r entsprechende Prothesenbasis Programm 4 U...

Страница 16: ...erden die vorgenom menen Einstellungen gel scht Das Ger t ist dann f r eine weitere Abscheidung bereit Galvanisierte Prothesenbasis entnehmen Schieben Sie nach Beendigung der Galvanisierung den Galvan...

Страница 17: ...rderen Durchf hrungen der Siliconabdichtung ziehen Nr 5 6 Kopf auf Halterung schieben Restgoldschwamm ausrichten Unterkante des Schwamms darf nicht unterhalb der Anode liegen Kopf auf Becherglas schie...

Страница 18: ...Schwamm verf rbt sich dunkel Durch Dr cken der Timer Taste werden die vorgenomme nen Einstellungen gel scht Das Ger t ist dann f r eine weitere Abscheidung bereit Nach Beendigung der Restgoldabscheidu...

Страница 19: ...en Restgoldanteil enthalten werden diese B der geschieden Scheidekosten auf Anfrage B der die bei uns ohne Lieferschein oder mit Liefer schein ohne entsprechenden Hinweis auf noch enthal tenes Gold ei...

Страница 20: ...der Eingabe der Stromstufen die maxi male F llmenge 550 ml berschritten ist eine weitere Erh hung der Stromstufe nicht mehr m glich Durch Dr cken einer beliebigen Strom stufentaste wird der entsprech...

Страница 21: ...beigegeben siehe Kapitel 2 l Die Goldbadmenge wurde falsch bemessen Gehen Sie bei der Berechnung und Abmes sung der Goldbadmenge genau nach der Ver gleichstabelle vor Goldbadmengenanzeige zur Kontroll...

Страница 22: ...pferstabes nicht mit Schrumpfschlauch berzogen ist ca 1 cm und sekrecht in den Kontakt eingef hrt wird l Das Objekt wurde in eine zu niedrige Stromstufe eingeordnet Die Oberfl che bestimmt die einzust...

Страница 23: ...uf dass der Gipsstumpf immer gen gend Abstand zum Becherglas hat Die Gipsst mpfe sind abgefallen l Es wurde zu wenig oder falscher Sekundenkleber verwendet Kontaktierungsbohrung zu gro Achten Sie dara...

Страница 24: ...ht entfernt werden Sie k nnen den Aufsatz bei Bedarf ersetzen Art Nr 6775 Reinigung der Anode mit Elektrodenreiniger J Art Nr 6692 Bel ge an der Anode k nnen die Goldabscheidung beeintr chtigen Um dem...

Страница 25: ...ch der Reinigung metallisch blank sein nicht mechanisch bearbeiten Bei starker Belagbildung ist die Einwirkzeit des Reini gers entsprechend zu verl ngern Nach Anwendung des Reinigers die Anode gr ndli...

Страница 26: ...5 C Genauigkeit 65 1 C 2 Temperaturf hler elektronisch geregelt Schutz Temperaturbegrenzer Sicherheitsabschaltung bei berschreitung der maximalen Heizplattentemperatur um mehr als 40 K durch zus tzlic...

Страница 27: ...nzzusatzmenge ml Chargen Nr 4 Becherglas klein mittel gro 5 Magnetr hrstab klein mittel gro 6 Datum der Abscheidung 1 Auftrags Nr Laufzeit min 2 Goldbadmenge ml Chargen Nr 3 Glanzzusatzmenge ml Charge...

Страница 28: ...28 1 10 0 K Ko on nf fo or rm mi it t t ts se er rk kl l r ru un ng g...

Страница 29: ...29 N No ot ti iz ze en n deutsch...

Страница 30: ...t system for manufacturing crowns anchor crowns for small bridges with ceramic or acrylic veneers as well as for mesostructures SupraCaps in implantology denture bases and secondary crowns in the tele...

Страница 31: ...process has been constantly monitored by leading universities Numerous independent publications have testified to the excellent clinical prognosis of restorations made with this process WIELAND DENTA...

Страница 32: ...6 2 Electroforming copings for crowns 8 3 Electroforming denture bases 15 4 Residual gold deposition 17 5 Recycling 19 6 Possible faults and troubleshooting 20 7 Cleaning instructions 24 8 Technical...

Страница 33: ...le damage Any damaged components must be serviced or exchanged by WIELAND DENTAL TECHNIK before the unit is put into operation l Check whether the line voltage indicated on the rating plate of the uni...

Страница 34: ...on for selecting the required power level can be adjusted individually for each channel LED display indicates the power level set for each channel Timer button for selecting the programs 1 electroform...

Страница 35: ...d upwards and remove it from its mount Ensure that the anode is inserted as far as it will go The AGC Micro enables up to six parts to be electro formed in a single process The parts are divided into...

Страница 36: ...le approx 1 mm Attach the AGC copper rod with a small amount of AGC contact adhesive Make sure that you remove any excess glue insulating layer If any other contact adhesive is used make sure this is...

Страница 37: ...capacities stated Determining the power level electroforming thick ness and quantity of gold electrolyte required Compare the size of the respective surfaces to be electroformed with the reference sam...

Страница 38: ...idges etc these can be contacted using several AGC copper rods In this case the power levels are added together Typical calculation 3 part splint e g 2 anterior teeth 1 premolar electroforming thickne...

Страница 39: ...sure effective electroforming and to obtain the maximum benefit from the gold electro lyte Please heed the shelf life of the gold electrolyte as well as any storage instructions Note that the gold ele...

Страница 40: ...any further The corresponding channel can be reset to zero by pressing any power level button Then remove the excess parts If you wish to obtain a thickness of less than 0 2 or 0 3 mm the next smalles...

Страница 41: ...t e g AGC copper rod not correctly inserted into the socket by pressing the Start Stop button Then press the Start Stop button once again The contacts will then be checked again If you do not interrup...

Страница 42: ...deleted by pressing the Start Stop key The unit must then be reset by pressing the timer button If an electroformed thickness of less than 150 m is achieved we advise against using the parts as they...

Страница 43: ...o the beaker and connect the top ends of the copper rods to the mount see sketch ensure that the denture base does not come into contact with the anode Press the timer button for the required denture...

Страница 44: ...erts to standby mode Press the timer button to clear the previous settings The unit is then ready for the next deposition operation Removing the denture base after electroforming After the electroform...

Страница 45: ...silicone seal and pull the copper rods through the two holes at the front of the silicone seal Nos 5 6 Push the head onto the mount position the residual gold sponge the lower part of the sponge must...

Страница 46: ...ng process The sponge takes on a darker colour Press the timer button to clear the previous settings The unit is then ready for the next deposition operation When the residual gold deposition process...

Страница 47: ...of worked out electrolyte Worked out AGC Micro gold electrolytes which contain some residual gold have to be refined Refining charges on demand Worked out electrolytes which are delivered without deli...

Страница 48: ...ty 550 ml is exceeded when entering the power level a further increase of the power level is not possible The corre sponding channel can be reset to zero by pres sing any power level button Then remov...

Страница 49: ...ume was incorrectly measured Calculate the precise gold electrolyte vol ume required on the basis of the comparison chart check by referring to the gold electrolyte quantity display This will help you...

Страница 50: ...e copper rod is not covered by shrink fit tubing leave approx 1 cm free l The part was assigned too low a position when correlating with the chart The surface determines the power level required Compa...

Страница 51: ...die and the beaker The plaster dies have fallen off l Either too little or the wrong contact adhesive was used or the contact holes were too big Ensure that the plaster die is securely fixed to the c...

Страница 52: ...otplate cover cannot be removed You can however exchange the cover if required Order No 6775 Cleaning the anode with electrode cleanser J Order No 6692 Deposits forming on the anode can be detrimental...

Страница 53: ...se After cleaning the anode should have a shiny metallic appearance do not polish mechanically If tough deposits have accumulated allow the cleanser more time to take effect After using the cleanser r...

Страница 54: ...lectrolyte 65 C 149 F tolerance 1 C 1 8 F 2 temperature sensor electronically controlled Safety temperature limiter additional relay activates a safety cut out if the maxi mum hotplate temperature is...

Страница 55: ...in 2 Electrolyte quantity ml LOT No 3 Brightener additive ml LOT No 4 Beaker small medium large 5 Magnetic agitator rod small medium large 6 Date of deposition 1 Order No Operating time min 2 Electrol...

Страница 56: ...28 N No ot te es s...

Страница 57: ...29 N No ot te es s english...

Страница 58: ...ur la confection d armatures pour couronnes couronnes d ancrage pour petits bridges avec recouvrement c ramique r sine ou m sostructures implantaires SupraCaps bases de proth ses totales et parties se...

Страница 59: ...ont par ailleurs fait tat des tr s bons pronos tics cliniques des restaurations r alis es avec ce proc d Avec son syst me AGC WIELAND DENTAL TECHNIK a t le premier fabricant au monde permettre la r al...

Страница 60: ...pour couronnes 8 3 Confection de bases lectroform es pour proth ses compl tes 15 4 Electrod position de l or r siduel 17 5 Recyclage 19 6 Probl mes ventuels et leurs solutions 20 7 Indications pour le...

Страница 61: ...ue l appareil n est pas ab m Si une pi ce de celui ci est endom mag e la faire r parer remplacer par WIELAND DENTAL TECHNIK avant de r utiliser l appareil l Afin d carter tout risque de destruction de...

Страница 62: ...t s lectionner le niveau d intensit du courant s lec tion du palier s par ment pour chaque canal Indicateur LED indique pour chaque connexion le palier s lectionn Touche timer elle sert s lectionner l...

Страница 63: ...r la t te de galvano vers le haut puis la d gager de son support Ins rer l anode quip e d une sonde de temp rature jusqu au point de but e L AGC Micro vous permet de r aliser jusqu 6 chapes la fois A...

Страница 64: ...e cuivre AGC dans le trou avec un peu de colle s chage rapide AGC Enlever ensuite toute la colle en trop car un exc s de colle constituerait une couche iso lante Si vous utilisez un autre type de coll...

Страница 65: ...paisseur de couche d lectrod position choisie est 0 2 mm 37 ml mais il faudra mettre 50 ml voir la page sui vante les quantit s de remplissage minimales respecter D finir les paliers d intensit de cou...

Страница 66: ...de brillance 10 Pour les pi ces lectrod poser de grande taille ensembles de 3 ou 4 couronnes par exemple vous pouvez placer plusieurs tiges de cuivre AGC Les paliers d intensit s additionnent Ensemble...

Страница 67: ...rrecte et une utilisation de l or la plus rentable possible Respecter la date limite de conservation de l lectro lyte or ainsi que les consignes de stockage indiqu es Attention Si de l lectrolyte or t...

Страница 68: ...ur de palier d intensit est alors impossible En pareil cas presser une touche palier quelconque pour remettre z ro le canal concern puis retirer les pi ces en trop Si l paisseur de la couche d lectrod...

Страница 69: ...ion par exemple tige de cuivre mal introduite dans son logement Apr s avoir solutionn le probl me r appuyer sur cette m me touche L appareil recontr le les connexions Si vous ne stoppez pas le process...

Страница 70: ...s Presser la touche marche arr t pour effacer le message d erreur puis r initialiser l ap pareil l aide de la touche timer Si l paisseur de la couche d lectrod position est inf rieure 0 15 mm nous vo...

Страница 71: ...connecter sur le support de la t te de galvano les tiges de cuivre comme indiqu sur le sch ma La base de proth se ne doit pas toucher l anode Presser la touche timer et s lectionner le programme qui...

Страница 72: ...min e l indicateur d tat affiche 000 la LED de fonctionnement s teint et l appareil se remet en mode standby Pour effacer les r glages effectu s presser la touche timer L appareil est alors pr t pour...

Страница 73: ...u s l avant trous n 5 et 6 Pousser la t te de galvano vers son support positionner l ponge le bord inf rieur de celle ci ne doit pas tre situ plus bas que l anode Ensuite placer la t te de galvano sur...

Страница 74: ...ste allum e L ponge prend une teinte plus fonc e Pour effacer les r glages effectu s presser la touche timer L appareil est alors pr t pour une nouvelle lectrod position Lorsque l lectrod position est...

Страница 75: ...sera r cup r par lectrod position tarif de l op ration de r cup ration nous consulter Les bains qui nous seront remis sans bon de livraison seront purement et simplement recycl s M me chose concernant...

Страница 76: ...e la quantit de remplissage maximale 550 ml est d pass e toute nouvelle augmentation d une valeur de palier est alors impossible Presser une touche palier quelconque pour remettre z ro le canal concer...

Страница 77: ...de brillance voir chapitre 2 l La quantit d lectrolyte or n tait pas correcte La d finir de mani re pr cise en se servant du tableau comparatif contr ler avec l indicateur de quantit de bain Ce tablea...

Страница 78: ...troduite bien verti calement dans la borne de connexion l La valeur du palier d intensit de courant s lec tionn e pour la pi ce tait trop faible La valeur de palier doit tre choisie en fonction de la...

Страница 79: ...fisamment loign du gobelet Les dies sont tomb s l Manque de colle ou colle inad quate ou trou du die trou pour la tige de cuivre trop grand Toujours faire en sorte que le die soit correcte ment coll l...

Страница 80: ...iens de maintenance agr s Les ventuelles t ches d lectrolyte or sur le couvercle de la plaque chauffante ne peuvent tre enlev es Si n cessaire remplacer le couvercle R f 6775 D capage de l anode avec...

Страница 81: ...e faire partir cause de l iode Apr s le nettoyage le m tal de l anode doit pr senter un aspect brillant ne pas la polir m caniquement S il y a beaucoup de d p ts sur l anode laisser le netto yant agir...

Страница 82: ...ature r gulation lectronique Limiteur de temp rature s curit Dispositif de mise l arr t automatique par relais en cas de d passement de la temp rature maximale de la plaque chauffante de plus de 40 K...

Страница 83: ...lot 4 Gobelet petit moyen grand 5 Agitateur magn tique petit moyen grand 6 Date de l lectrod position 1 N de commande Dur e min 2 Quantit d lectrolyte ml N de lot 3 Quantit d additif de brillance ml N...

Страница 84: ...28 N No ot te es s...

Страница 85: ...29 N No ot te es s fran ais...

Страница 86: ...a realizzazione di corone corone d ancoraggio per piccoli ponti con rivestimenti in ceramica o resina mesostrutture SupraCaps nella protesica degli impianti corone secondarie nella tecnica delle coron...

Страница 87: ...accompa gnato scientificamente dalle pi famose universit Numerose pubblicazioni indipendenti documen tano l ottima prognosi clinica di restaurazioni realizzate con questo procedimento Con il sistema...

Страница 88: ...cchiatura 6 2 Galvanizzare corone 8 3 Galvanizzare basi di protesi 15 4 Elettro deposizione dell oro residuo 17 5 Recycling 19 6 Fonti di errore 20 7 Indicazioni sulla pulizia 24 8 Dati tecnici 26 9 P...

Страница 89: ...All interno non contiene parti che richiedino manutenzione da parte dell utilizzatore Badare che nessun oggetto giunga attraverso le feritoie di ventilazione all inter no dell apparecchiatura Questo p...

Страница 90: ...unzionamento Tasti ed indicazioni Tasto livello di tasto per impostare il livello di corrente corrente necessario regolabile singolamente per ogni canale Indicazione LED indica il livello di corrente...

Страница 91: ...anizzazione verso l alto ed estrarla dal supporto Inserire l anodo con la sonda termica facente parte della fornitura fino all arresto Con la AGC Micro si possono produrre fino a sei elementi galvaniz...

Страница 92: ...mm Incollare la bacchetta in rame AGC con poca colla istanta nea cianocrilato AGC Togliere assolutamente la colla in eccesso strato isolante Usando una colla istantanea diversa da quella consigliata...

Страница 93: ...Spessore 0 2 mm 49 ml 28 ml 77 ml Elemento con livello di corrente 4 Spessore 0 2 mm 37 ml impiego 50 ml vedi quantit di riempimento indicata Stabilire livello di corrente spessore e quantit di bagno...

Страница 94: ...3 o 4 corone questi possono essere contattati con pi bacchette di rame AGC Le correnti si sommano quindi Esempio di calcolo Bloccaggio di 3 corone p e 2 denti frontali 1 premolare Spessore 0 2 mm 86...

Страница 95: ...zione affidabile ed uno sfruttamento massimo del contenuto d oro Osservare la data di scadenza dei bagni d oro e le indi cazioni sul magazzinaggio Attenzione il bagno d oro causa macchie sull appa rec...

Страница 96: ...n pi possibile Premendo uno qualsiasi dei tasti di livello corrente il rispettivo canale viene riportato a zero Togliere quindi gli elementi in soprannumero Se si desidera ottenere uno spessore inferi...

Страница 97: ...asto Start Stop potete interrompere il processo ed eliminare l errore di contatto p e bacchetta di rame AGC non ben inserita nella boccola Successivamente premere di nuovo il tasto Start Stop Ora vien...

Страница 98: ...8 Le regolazioni vengono mantenute Tra mite il tasto Start Stop viene cancellato l avviso di errore Successivamente tramite i tasto Timer si deve ripristinare l apparecchiatura Se lo spessore de gli e...

Страница 99: ...rsi al di sotto dell anodo Spingere la testa al di sopra del bicchiere e contattare i fili di rame sul supporto vedi schizzo Badare che la base di protesi non appoggi contro l anodo Premere il tasto T...

Страница 100: ...ngono cancellate le rego lazioni impostate L apparecchiatura quindi pronta per un ulteriore galvanizzazione Prelievo della base di protesi galvanizzata Terminata la galvanizzazione spingere la testa d...

Страница 101: ...in sili cone e tirare i fili di rame attraverso i due passanti anteriori della guarnizione di silicone n 5 6 Spin gere la testa sul supporto orientare la spugna per l oro residuo il bordo inferiore d...

Страница 102: ...curo Premendo il tasto Timer le regolazioni impostate ven gono cancellate L appareccchiatura quindi pronta per un ulteriore elettrodeposizione Terminata l elettrodeposizione dell oro residuo indica zi...

Страница 103: ...iduo i bagni non contengono pi metallo prezioso Per lo smaltimento dei bagni consumati dopo l elettrodeposizione dell oro residuo preghiamo di mettersi in contatto con il proprio Concessionario 5 5 R...

Страница 104: ...quantit massima di riempimento 550 ml non pi possibile un ulteriore aumen to del livello di corrente Premendo uno qualsiasi dei tasti di livello corrente il rispettivo contatto viene riportato a zero...

Страница 105: ...bagno d oro stata calcolata in modo sbagliato Procedere per il calcolo e la misurazione del bagno d oro esattamente secondo la tabella di confronto indicazione della quantit di bagno d oro per contro...

Страница 106: ...ta di tubicino termoretraibile ca 1 cm e che sia inserito per pendicolarmente nel contatto l L elemento stato classificato secondo un livello di corrente troppo basso La superficie da galvanizzare det...

Страница 107: ...oncone di gesso abbia sempre sufficiente distanza dal bicchiere I monconi di gesso si sono staccati l stata usata troppo poca colla istantanea o un tipo sbagliato e o il foro di contattamento era trop...

Страница 108: ...gno d oro sul sopralzo della piastra di riscaldamento non possono essere tolte All occorrenze si pu sostituire il sopralzo Art n 6775 Pulizia dell anodo con detergente per elettrodi J Art n 6692 Patin...

Страница 109: ...ulizia l anodo dovrebbe apparire lucido metallico non trattarlo meccanicamente In caso di forte formazione di patina prolungare corris pondentemente il tempo di azione del detergente Dopo l impiego de...

Страница 110: ...e Limitatore di protezione della temperatura disinserzione di sicurezza tramite rel supplementare al superamento di oltre 40 K della temperatura di ris caldamento massima della piastra di riscaldament...

Страница 111: ...e ml Lotto n 4 Bicchiere piccolo medio grande 5 Barra agitatrice magnetica piccola media grande 6 Data galvanizzazione 1 Ordine n Durata min 2 Quantit di bagno d oro ml Lotto n 3 Quantit di additivo l...

Страница 112: ...28 N No ot ta a...

Страница 113: ...099 Dresden Germany Fon 49 3 51 81 70 30 Fax 49 3 51 81 70 35 5 WIELAND Dental Technik GmbH Co KG Niederlassung D sseldorf Sternstra e 28 30 40479 D sseldorf Germany Fon 49 211 4 91 96 97 0 Fax 49 211...

Отзывы: