CAUTION
Do not use excessive force when releasing pressure on
the connector. Excessive force may result in breaking the sliding
outer piece.
To remove a ZIF connector
CAUTION
Remove a flex cable only
after
the ZIF latch is open.
1. Using a suitable tool (for example, a paper clip, small flat-head screwdriver, or needle-nose
pliers), slide the latching piece of the connector away from the connector body.
2. Remove the cable.
To connect a ZIF connector
1. Slide the latching piece of the connector away from the connector body.
2. Insert the flex cable into the connector. This may require using a suitable tool to keep the
latching piece elevated.
3. Slide the latching piece toward the connector body until it locks into place.
•
FHY connector (Self-Lock mechanism)
: The device uses Self-Lock mechanism cable
connectors.
A Self-Lock mechanism enables the cover to close by one-touch operation.
34 Disassembly and repair