Attach the pump assembly to the main PCA board
Kit item:
•
105920, PUMP ASSEMBLY AND 2 T-10 TORX SCREWS
1. Place the pump manifold onto the main PCA board transducer. Ensure that the pump
manifold is fully seated on the transducer.
2. Connect the pump assembly connector to the main PCA board.
3. Turn over the main PCA board.
4. Secure the pump assembly to the main PCA board with 2 T-10 Torx screws.
Replace the main PCA board
Replace the main PCA board as needed after removing all other components. Before you begin,
ensure that you have the appropriate replacement kit items:
•
107184, SRV KIT, MAIN PCA, AND LCD FRAME
NOTE
After installing a new PCA board, use the Welch Allyn Service Tool, Gold
Edition, to prepare the board for first-time use. This includes programming the
device serial number into the new board and performing an NIBP initialization
during full device functionality. See "Program the main PCA board" and "Initialize
the NIBP board" for more information. After replacing the main PCA board and
completing the full functional check, set the first boot option to Production
Complete using the service tool. This will set the device to prompt you to enter
the language, date and time on first boot. See "
Service Tool Spot 4400 Help File"
for
more information.
Service manual
Disassembly and repair 55
Содержание Spot Vital Signs 4400
Страница 1: ...Welch Allyn Spot Vital Signs 4400 Service manual Software version 1 X...
Страница 8: ...4 Symbols and definitions...
Страница 34: ...30 Power up sequence...
Страница 68: ...64 Field replaceable units...
Страница 90: ...Material no Item 4400REPRC WEB TRN 4400 Repair re certification web training 86 Options and upgrades...
Страница 127: ...Interconnect diagram Service manual Appendix 123...
Страница 128: ...124 Appendix...
Страница 132: ......