Message
Possible cause
Suggested action
Invalid algorithm. Select correct
algorithm and retry. 050512
Invalid NIBP Algorithm. NIBP
component software tried to
configure the sensor in an illegal
manner
Verify the algorithm. If the problem
persists, have the device inspected
by a qualified service technician.
NIBP not functional. 050513
Invalid NIBP initiation code
Internal malfunction. If the problem
persists, have the device inspected
by a qualified service technician.
Invalid patient mode. Select correct
patient mode and retry. 050514
Invalid patient mode on NIBP. The
NIBP component software tried to
configure the sensor in an illegal
manner
Verify correct patient mode. If the
problem persists, have the device
inspected by a qualified service
technician.
NIBP not functional. 050515
Invalid module configuration for
NIBP
Internal malfunction. If the problem
persists, have the device inspected
by a qualified service technician.
NIBP not functional. 050516
NIBP module malfunction
Internal malfunction. If the problem
persists, have the device inspected
by a qualified service technician.
Ambient temperature out of range.
Clear error and retry. 050517
Ambient temp out of range on
NIBP
Restore device to normal
temperature ranges and try again.
Low battery. Plug into outlet.
050518
The NIBP module power rail is too
low
Plug device into an AC outlet to
charge the battery.
Battery overcharged. Disconnect
from outlet. 050519
The NIBP module power rail is too
high. NIBP was calibrated without
A/C connected or battery too low.
Battery is overcharged. Remove
from charging source. Charge
battery and initialize NIBP module
and recalibrate with A/C
connected.
NIBP not calibrated. Calibrate the
module. 050601
NIBP failed to load safety processors
calibration record from EEPROM
Calibrate the NIBP module. If the
problem persists, have the device
inspected by a qualified service
technician.
NIBP not functional. 050602
NIBP safety processor failed ROM
checksum
Internal malfunction. If the problem
persists, have the device inspected
by a qualified service technician.
NIBP not calibrated. Calibrate the
module. 050603
NIBP safety processor not
calibrated, missing calibration
signature
Calibrate the NIBP module. If the
problem persists, have the device
inspected by a qualified service
technician.
Cuff pressure limits exceeded.
050604
NIBP system failure. Overpressure
Restrict patient movement.
Premature auto cycle skipped.
050605
NIBP auto cycle skipped, SVRP
requirement not satisfied
Cuff pressure is not below safety
return pressure long enough to
allow a cycle to occur.
Cuff pressure too high. Clear error
to retry. 050606
NIBP cuff pressure above SVRP for
too long
Verify cuff connections. If the
problem persists, have the device
Service manual
Appendix 107
Содержание Spot Vital Signs 4400
Страница 1: ...Welch Allyn Spot Vital Signs 4400 Service manual Software version 1 X...
Страница 8: ...4 Symbols and definitions...
Страница 34: ...30 Power up sequence...
Страница 68: ...64 Field replaceable units...
Страница 90: ...Material no Item 4400REPRC WEB TRN 4400 Repair re certification web training 86 Options and upgrades...
Страница 127: ...Interconnect diagram Service manual Appendix 123...
Страница 128: ...124 Appendix...
Страница 132: ......