System messages
Message
Possible cause
Suggested action
000001
System failure
Restart the device
000002
System failure
Restart the device
000003
System failure
Restart the device
000004
System failure
Restart the device
000005
System failure
Restart the device
000006
System failure
Restart the device
Internal hardware failure. The root file system is corrupted;
restart not possible
Restart the device. If the problem persists,
replace the main PCBA.
Internal hardware failure.
140100
EEPROM access failed. The device
boot is possible, but wired
communications are disabled
Reprogram the EEPROM. If the problem
persists replace the main PCBA.
Internal hardware failure. SPL memory test failure, the device
will sound an SOS pattern
Restart the device. If the problem persists,
replace the main PCBA.
Internal hardware failure.
1C1000
The device PIC communications
never starts or quits. The
communication won’t reasonably
recover at startup or during
operation
Restart the device. If the problem persists,
replace the main PCBA.
Low battery 30 minutes
or less remaining. 1C1005
The battery power is low
Connect the power supply to AC power to
charge the device.
Low battery 5 minutes or
less remaining. 1C1006
The battery power is extremely low Connect the power supply to AC power to
charge the device.
Battery is critically low;
plug into outlet. Device is
shutting down. 1C1007
The battery power is critically low
Connect the power supply to AC power to
charge the device.
Update unsuccessful.
Reboot and retry. 1C1008
The software update failed
Restart the device. If the problem is still
present, replace the main PCBA.
Host battery not
charging. 1C100A
The host battery is not charging
Restart the device. If the problem is still
present, check cable connections. If the
problem is still present, perform functional
checks for host. If the problem is still present,
replace battery. If the problem persists,
replace the main PCBA.
Factory default settings
now active. 3A0001
The factory configuration settings
are active
The device has been configured to factory
defaults, any user settings have been reset.
Internal hardware failure.
Device will shut down.
1C100D
Power supply issue. The PMIC is too
hot
Check the operating environment
temperature. Allow the device to cool down
before operating again. If the problem is still
present, check the cable connections. If the
problem is still present, perform functional
118 Appendix
Содержание Spot Vital Signs 4400
Страница 1: ...Welch Allyn Spot Vital Signs 4400 Service manual Software version 1 X...
Страница 8: ...4 Symbols and definitions...
Страница 34: ...30 Power up sequence...
Страница 68: ...64 Field replaceable units...
Страница 90: ...Material no Item 4400REPRC WEB TRN 4400 Repair re certification web training 86 Options and upgrades...
Страница 127: ...Interconnect diagram Service manual Appendix 123...
Страница 128: ...124 Appendix...
Страница 132: ......