Message
Possible cause
Suggested action
NIBP not functional. 050105
WACP message CRC mismatch on
NIBP module
Internal malfunction. If the problem
persists, have the device inspected
by a qualified service technician.
NIBP not functional. 050201
This message is not implemented
by the NIBP module
Internal malfunction. If the problem
persists, have the device inspected
by a qualified service technician.
NIBP not functional. 050202
This message is not supported by
the NIBP module
Internal malfunction. If the problem
persists, have the device inspected
by a qualified service technician.
NIBP not functional. 050203
The NIBP module has run out of
memory
Internal malfunction. If the problem
persists, have the device inspected
by a qualified service technician.
NIBP not functional. 050205
The NIBP module has received a
invalid parameter
Internal malfunction. If the problem
persists, have the device inspected
by a qualified service technician.
NIBP not functional. 050206
The parameter provided by the
NIBP module is outside of the
allowable range for the specified
message
Internal malfunction. If the problem
persists, have the device inspected
by a qualified service technician.
NIBP not functional. 050207
The NIBP module message requires
an object, but did not contain one
Internal malfunction. If the problem
persists, have the device inspected
by a qualified service technician.
NIBP not functional. 050208
The NIBP module object provided
with the message could not be
deserialized
Internal malfunction. If the problem
persists, have the device inspected
by a qualified service technician.
NIBP not functional. 050209
The NIBP module object could not
be serialized
Internal malfunction. If the problem
persists, have the device inspected
by a qualified service technician.
NIBP not functional. 05020A
The NIBP module message is
performing a request or action
when the module state prohibits
the request or action.
Internal malfunction. If the problem
persists, have the device inspected
by a qualified service technician.
NIBP not calibrated. 050503
Factory EEPROM checksum error on
NIBP. Units internal configuration
was corrupted
Internal malfunction. If the problem
persists, have the device inspected
by a qualified service technician.
NIBP not functional. 050504
User EEPROM checksum error.
Configuration data which can be
set in the user's configuration
menu was damaged or lost on NIBP
Calibrate the NIBP Module. If
problem persists, have the device
inspected by a qualified service
technician.
NIBP not functional. 050505
Post failure of A/D convertor
Internal malfunction. If the problem
persists, have the device inspected
by a qualified service technician.
NIBP not calibrated. Calibrate the
module. 050509
NIBP module calibration failure, the
calibration signature is zero
Calibrate the NIBP module.
106 Appendix
Содержание Spot Vital Signs 4400
Страница 1: ...Welch Allyn Spot Vital Signs 4400 Service manual Software version 1 X...
Страница 8: ...4 Symbols and definitions...
Страница 34: ...30 Power up sequence...
Страница 68: ...64 Field replaceable units...
Страница 90: ...Material no Item 4400REPRC WEB TRN 4400 Repair re certification web training 86 Options and upgrades...
Страница 127: ...Interconnect diagram Service manual Appendix 123...
Страница 128: ...124 Appendix...
Страница 132: ......