Message
Possible cause
Suggested action
persists, have the device inspected by a
qualified service technician.
Replace temperature probe. 030808 The temperature module
probe was not
characterized/calibrated
Probe malfunction. Replace the probe. If the
problem persists, have the device inspected
by a qualified service technician.
Insert correct color-coded probe
well. 030809
The temperature module
is missing the probe well
Insert the probe well
Temperature not functional. 03080A The temperature module
has a problem saving to
the device EEPROM in
biotech mode
Internal malfunction. If the problem persists,
have the device inspected by a qualified
service technician.
Temperature not functional. 03080B The temperature module
error detection
mechanism detected an
error
Internal malfunction. If the problem persists,
have the device inspected by a qualified
service technician.
Replace temperature probe. 03080C The temperature module
probe error detection
mechanism detected an
error
Probe malfunction. Replace probe. If the
problem persists, have the device inspected
by a qualified service technician.
Temperature not functional.
03080D
The temperature module
log error detection
mechanism detected an
error
Probe malfunction. Replace probe. If the
problem persists, have the device inspected
by a qualified service technician.
Temperature not functional. 03080E The temperature module
calibration error
detection mechanism
detected an error
Probe malfunction. Replace probe. If the
problem persists, have the device inspected
by a qualified service technician.
Connect temperature probe.
03080F
The temperature module
detected no probe
connected
Probe malfunction. Replace probe. If the
problem persists, have the device inspected
by a qualified service technician.
Replace temperature probe. 030810 The temperature module
cannot read the probe
EEPROM correctly or the
probe left the factory
without being tested.
Probe malfunction. Replace probe. If the
problem persists, have the device inspected
by a qualified service technician.
Temperature not functional. 030811 The temperature module
has an invalid events
index
Internal malfunction. If the problem persists,
have the device inspected by a qualified
service technician.
Temperature not functional. 030812 There is a problem
reading the temperature
module EEPROM or
saving to the device
EEPROM in biotech
mode.
Internal malfunction. If the problem persists,
have the device inspected by a qualified
service technician.
Replace temperature probe. Code
030813
The temperature module
has a problem reading
the probe EEPROM.
Probe malfunction. Replace probe. If the
problem persists, have the device inspected
by a qualified service technician.
114 Appendix
Содержание Spot Vital Signs 4400
Страница 1: ...Welch Allyn Spot Vital Signs 4400 Service manual Software version 1 X...
Страница 8: ...4 Symbols and definitions...
Страница 34: ...30 Power up sequence...
Страница 68: ...64 Field replaceable units...
Страница 90: ...Material no Item 4400REPRC WEB TRN 4400 Repair re certification web training 86 Options and upgrades...
Страница 127: ...Interconnect diagram Service manual Appendix 123...
Страница 128: ...124 Appendix...
Страница 132: ......