Page 29
990008 Rev. H
•
Tubing pinched
•
Mixing Chamber empty
•
Air Valve problem
•
Clogged orifice in Chamber
panel
•
Check tubing
•
Refill Mixing Chamber
with powder
•
Call local service rep
•
Disassemble and clean
Chamber
Air flow from nozzle
when footswitch is not
depressed
•
Loss of air pressure due to a
misadjusted air valve
regulator
•
Faulty footswitch
•
Pinch Valve block is not
tight
•
Call local service rep
•
Call local service rep
•
Tighten thumb screws
Erratic powder flow
•
Powder caking in chamber
•
Chamber empty
•
Nozzle partially clogged
•
Vibrator inoperative
•
Moisture in air line
•
Blown pinch tube
•
Replace powder
•
Check powder level
•
Check nozzle
•
Call local service rep
•
Check filters, replace
powder
•
Change pinch tubes
Powder spray loses
definition
•
Worn nozzle tip
•
Too much distance from
cutting surface
•
Replace nozzle
•
Move closer
Excessive powder inside
of unit
•
Hole in tubing.
•
Hole in pinch tube
•
Exhaust bottle filter loose
•
Replace tubing
•
Replace pinch tube
•
Tighten or replace
filter
Remote error light turns
on when foot switch is
pressed
•
Remote handpiece micro
switch was not pressed
•
Toggle micro switch