Page 22
990008 Rev. H
•
Lock both doors, removing key.
CAUTION: Make certain that no lint, dirt, or other foreign substances enter
the chamber. This can result in plugging the unit. Failure to turn the power off
will prevent removal of the mixing chamber lid.
Operational Checkout
•
Place the Power Switch to ON and make sure the Particle Energy indicator is set to
the desired setting. If not, change it by touching the up or down arrow accordingly.
A Particle Energy setting of 80 is suggested when first using the instrument.
•
Set the Beam Intensity to the desired setting, adjusting it up or down by touching the
appropriate arrow. A suggested beginning setting is 7.0 for the Beam Intensity.
•
Touch the Air control to ON.
•
•
Lift the handpiece from the cradle.
If the handpiece is not in the cradle, the
microswitch on top of the Remote must be depressed before the footswitch
is activated.
•
•
Press the footswitch to obtain powder flow through the nozzle.
You can now practice operative procedures on extracted teeth as described in
Chapter 2, Safety.
Footswitch Operation
•
Depressing the footswitch enables the powder to flow through the handpiece/nozzle.
Unit Shutdown
•
Set the Air Switch to the Standby position.
•
Set the Power Switch to the OFF position.
Mixing Chamber Powder Unloading
•
Follow unit shutdown instructions above.
•
Unlock the left door and open.
•
Loosen the black knob holding the block in place. The block will rotate down,
releasing the mixing chamber so it can be removed for emptying.
•
Firmly hold the chamber and loosen the mixing chamber lid by turning it
counterclockwise.