Page 6
990008 Rev. H
Full Diagnostic
•
At start-up
•
Continuous monitoring
Additional Features
•
Quickly detachable handpiece for easy sterilization
•
Double doors for simplifying preventive maintenance
•
Exhaust bottle container
•
Easy pinch tube replacement
•
Easy access to powder canister
•
Accessory drawer for nozzles, handpieces, etc.
•
Full range of operation from 30 psi to 120 psi
•
Convenient handles
•
Mobility
•
Wide range of nozzles in sizes/angles and cutting capabilities
•
Locked doors to prevent unauthorized access
Specifications
•
Power Requirements:
Domestic
Mach 5.0 100-127 V~, 60 Hz, 400 mA
Mach 5.0 PLUS 100-127 V~, 60 Hz, 6.5 A
International
Mach 5.0 220-240 V~, 50 Hz, 170 mA
Mach 5.0 PLUS 220-240 V~, 50 Hz, 3.5 A
NOTE: Verify unit has correct voltage for your
application.
•
Air Requirements:
Mach 5.0, 80 to125 psi (550 to 860 kPa) with
at least 1.5 cfm (42.5 l/min) recommended; the
supply pressure must be at least 5 psi (35 kPa)
greater than the selected operating pressure.
The Mach 5.0 PLUS incorporates a built-in
compressor which has an automatic thermal
Содержание Kreativ Mach 5.0
Страница 13: ...Page 13 990008 Rev H Drawings of Controls Figure 1...
Страница 14: ...Page 14 990008 Rev H Figure 2...
Страница 16: ...Page 16 990008 Rev H...
Страница 35: ...Page 35 990008 Rev H THERE ARE NO HAZARDOUS MATERIALS PRESENT...