GRANUDOS Touch, control unit
Index: 02
Date modified: 11/06/2019
OI No.: BA SW 010-02 Granudos Touch-Steuereinheit EN SCHW 8.docx
Page 23 of 43
Main Menu
Service
Input Test
The input test serves for checking the connected inputs (switches). The
changing activation of the switch inputs is indicated by 0 (open) or 1
(closed).
The fourth column displays the pin header (SLx) and the connectors
(Pinx/x) to which the switch is connected.
The fifth column shows the function of the switches NO or NC,
respectively.
NO (normally open) indicates open in the operating state and closed in
the event of a fault.
NC (normally closed) indicates closed in the operating state and opened
in the event of a fault.
You can scroll through the 4 pages using the + and – buttons.
Main Menu
Service
Output test
The output test is used for checking the connected outputs (pumps,
motors and relays). The selected output is activated for 30 seconds.
The control can be deactivated at any time using “Stop”.
For safety reasons (generation of chlorine gas), the output test for the
chemical-dosing outputs is only released if no fault exists that could
prevent the dosing.
An output test can be performed for the following actuators:
pH dosing
Chlorine dosing
Knocker
Solenoid valve buffer tank
Alarm
Chemical reserve
Chemical empty
Dust extraction flushing
Main Menu
Service
Operation hour counter
Reset operating hours for each individual actuator.