background image

GRANUDOS Touch, control unit

 

 

 

Index: 02 

Date modified: 11/06/2019 

OI No.: BA SW 010-02 Granudos Touch-Steuereinheit EN SCHW 8.docx 

Page 16 of 43 

 

Main Menu 

 Settings 

 dosing performance chlorine/acid 

The  dosing  performance  menu  serves  for  adjusting  the  dosing 
performance to the expected consumption of chemicals in the pool.  

Especially in case of lower water content, it is very important to adjust 
the dosing performance.  
The required dosing performance depends on several factors, e.g., the 
basin volume, location, type of use and of course the frequency of use 

by pool visitors. 
 
 
 

 

Explanation of setting parameters 

 

Chlorine:  

  Dosing time of chlorine granulate adjustable from 1-15 seconds 

 

Acid:  

  Dosing time of acid in seconds adjustable from 1-8 seconds 

 

 

ATTENTION!   
The use of sulphuric acid is generally possible up to a concentration of 50%. For higher 

concentrations  or  when  using  other  acids  (e.g.  hydrochloric  acid,  dissolved  sodium 
bisulphate, etc.), the changed dosing performance and/or increased corrosiveness must 

be observed! We recommend to consult with the company WDT!  

 

 

Cycle:  

 Length of complete dosing cycle adjustable from 30-360 seconds 

 

Dosing time limit acid:  

  Dosing time monitoring for external control adjustable from 0-100 minutes  

 

Dosing time limit chlorine  

  dosing time monitoring for external control adjustable from 0-100 minutes 

 

Manual dosing off / on:  

  Switch between automatic and manual dosing 

 

Explanations of the dosing process 

Chlorine and acid are dosed at intervals with pauses between the dosings. 
The dosing is organised in dosing cycles (time from dosing interval to dosing interval) and dosing times (dosing times for 
chlorine and acid dosing motors).  
 

A dosing cycle proceeds like this: 
1.

 

Chlorine dossing - 1-15 sec.  

2.

 

Pause - 3.5 sec.  

3.

 

Acid dosing - 1-8 sec.  

4.

 

Pause until end of cycle 

 
This cycle runs continuously with "manual dosing".   
In "Automatic" mode, the cycle will also run, but the dosing will only be activated if the control command for dosing (chlorine 

or acid) is pending. If the control command hits for a pause, the corresponding dosing will be activated in the next dosing 
cycle. 
 
Example for setting the dosing performance 

Prerequisite: Indoor pool with a circulation capacity of 200m³/h. In accordance with DIN 19643, in indoor pools 2g of 
chlorine must be added per each 1m³/h circulation capacity. 
Calculation:  200m³/h x 2g chlorine = 400g/h chlorine dosing 
Maximum dosing performance 

GRANUDOS 45

 = 2000g/h = 100% 

That means that the dosing performance must be set to 20% for a chlorine dosing of 400g/h. 
The acid dosing is initially set to the same value, i.e., at 20%. 

A readjustment may be necessary during ongoing operation. 

 

Explanation of dosing time monitoring  
If the GRANUDOS is controlled by a measuring and control device, the dosing performance must be set sufficiently high so 

that even large loads can be regulated without large deviations from the target value. This means that the pause times must 
always be greater than the dosing times. If, however, the dosing times are longer than the pause times, there is a fault with 
the measuring and control device (e.g. control relay hanging) or the dosing performance is too low (set too weak or motor 
or dosing screw faulty) or the input relay on the GRANUDOS control plate hangs. The dosing time monitoring totals the 

dosing times against the pause times and if the set dosing time limit is exceeded, dosing is deactivated and indicated as a fault. 

 

Both the chlorine-control and pH-control or -dosing is monitored. 

 

Содержание Granudos 10 Touch

Страница 1: ...d Installation instructions Control Unit Type GRANUDOS Touch Operating instructions Part 2 for control unit Touch Part 2 for operating instructions for dosing device Granudos 10 Touch No OI SW 002 Gra...

Страница 2: ...ation display Automatic operation 10 Start Delay booster pump and dosing delay 13 Automatic operation 13 Alarm 14 Manual dosing emergency programme 14 6 3 The Main Menu 14 Main Menu Login 15 Main Menu...

Страница 3: ...d translation into other languages even in excerpts are only permitted with the express authorisation of the company WDT Werner Dosiertechnik GmbH Co KG These operating instructions are an English tra...

Страница 4: ...eral notices DANGER DANGER denotes a safety notice which if disregarded may lead to serious or life threatening injuries or serious material damage CAUTION CAREFUL denotes a safety notice which if dis...

Страница 5: ...r subsequent damages and for the processing of warranty claims are excluded There are no warranty claims for damages caused by frost water and electrical overvoltage or by improper handling Tip In ord...

Страница 6: ...ch Emergency Poison Centre or any other Poison Centre Phone 49 89 19240 Excerpt from the Accident Prevention Regulations GUV V D 5 Installation rooms for chlorination systems and storage rooms Section...

Страница 7: ...is used for the following 2 dosing devices GRANUDOS 10 Touch GRANUDOS 45 100 Touch The control system has the following main functions dosing chlorine granulate with screw conveyor Dosing acid with ho...

Страница 8: ...22VA safety Schuko plug Protection class Casing IP54 Casing IP54 Interface connection Modbus TCP USB for data export Modbus TCP USB for data export Operating data Measuring range pH value 2 00 to 12 0...

Страница 9: ...list personnel from a specialist company Prior to commissioning the installed devices must be checked for proper installation and leaks The device has been delivered with specific factory settings The...

Страница 10: ...gure 2 Automatic operation In automatic operation the current operating status fill levels and active inputs and outputs are displayed IN OUT The inputs and outputs can be operation messages or fault...

Страница 11: ...gulation empty yellow disinfection level container for disinfectant empty blue flocculation level container for flocculant empty High chlorination e g for filter disinfection active The controller is...

Страница 12: ...The following symbols relate only to the buffer filling OPTION IN List optional design with buffer tank Buffer tank filling is starting Buffer tank filling is ending The buffer tank level has dropped...

Страница 13: ...alarms are suppressed during this time The device then switches to automatic operation In the event of a booster pump delay the indicated delay runs out and the booster pump then starts Serves to vent...

Страница 14: ...n alarm will be automatically deleted when its cause has been remedied When for example the empty chemical container has been replaced However the alarm switch off time monitoring must be acknowledged...

Страница 15: ...The settings menu is used to implement the desired settings for the dosing device Using the white arrow keys on the side you can navigate to the next screen menu The 2nd page is only displayed in user...

Страница 16: ...orine and acid dosing motors A dosing cycle proceeds like this 1 Chlorine dossing 1 15 sec 2 Pause 3 5 sec 3 Acid dosing 1 8 sec 4 Pause until end of cycle This cycle runs continuously with manual dos...

Страница 17: ...the GRANUDOS to avoid acid overdosing in the event of a fault in the chlorine dosing Manual dosing Off On It is possible to change over to continuous manual dosing This may be required in the event of...

Страница 18: ...he concentration must be analysed photometrically during shock chlorination Main menu Settings System Date Time Set date and time Password Assign a password Display brightness Adjust the display brigh...

Страница 19: ...sheet If you place a green check mark next to Clear text the entered numbers are shown instead of white dots To change an active end user password it must be entered in the upper line The new passwor...

Страница 20: ...erator must establish the necessary IT requirements for remote access e g VNP connection data security etc e Main menu Settings System Reset This order can only be performed in user level 2 All of the...

Страница 21: ...so that the required concentration is present after the buffer tank is filled For monitoring purposes the concentration and pH value must be analysed repeatedly after the buffer tank has been filled T...

Страница 22: ...en installed in the dosing device Otherwise the dosing quantities will be determined incorrectly Main Menu Settings Dosing performance acid default fit in dosing components These settings can only be...

Страница 23: ...y closed indicates closed in the operating state and opened in the event of a fault You can scroll through the 4 pages using the and buttons Main Menu Service Output test The output test is used for c...

Страница 24: ...confirmed individually The previous operating day s consumption left column cannot be reset Main Menu Service Maintenance interval The time interval for the maintenance message in days is set here Af...

Страница 25: ...u select another day by touching it that day will be shown on a white background You can use the Event Log ICON to view the events of the selected day b Main menu Log Data export You can use the Expor...

Страница 26: ...suring amplifier pH7 buffer solution for determining offset voltage The optimal offset voltage is at 0mV up to 30mV pH4 second buffer solution for determining conductance voltage Slope Inclination mV...

Страница 27: ...ly provide incorrect values Carry out the calibration according to the instructions Please observe the displayed notices and follow the menu navigation The set deviation is displayed as offset at the...

Страница 28: ...e cleaned using the electrode cleaner supplied with the delivery For this purpose the glass shaft of the pH electrode is submerged in the electrode cleaner for a few minutes Depending on the state and...

Страница 29: ...er lid lugs run in guide grooves up to the front lock in position Afterwards the display lid is raised The display lid can be supported on the main casing using a support bracket so that it remains ra...

Страница 30: ...hing tank level maximum ALARM The water level in the flushing tank is too high less water is being siphoned off than is flowing into the flushing tank through the float valve Dosing is being stopped T...

Страница 31: ...osing 15 Buffer tank filling timeout ALARM Filling occurred too slowly Buffer tank filling has been deactivated by the system 1 Check switching valves 2 Check Start buffer tank sensor Its contact must...

Страница 32: ...filled to about half way with electrolyte and pushed onto the glass shaft of the electrode The glass electrodes are frost resistant to approx 15 C if the temperature falls below this value the electro...

Страница 33: ...No BA SW 010 02 Granudos Touch Steuereinheit EN SCHW 8 docx Page 33 of 43 9 Documents 9 1 Declaration of conformity See operating instructions Part 1 regarding the respective dosing device for Granud...

Страница 34: ...grams DANGER Risk of death due to high voltage All electrical work on the device must only be carried out by trained specialists in accordance with the applicable safety regulations Internal fuses mus...

Страница 35: ...GRANUDOS Touch control unit Index 02 Date modified 11 06 2019 OI No BA SW 010 02 Granudos Touch Steuereinheit EN SCHW 8 docx Page 35 of 43 Wiring diagram I O board GRANUDOS 10...

Страница 36: ...GRANUDOS Touch control unit Index 02 Date modified 11 06 2019 OI No BA SW 010 02 Granudos Touch Steuereinheit EN SCHW 8 docx Page 36 of 43 Wiring diagramm Buffer tank GRANUDOS 45 100...

Страница 37: ...GRANUDOS Touch control unit Index 02 Date modified 11 06 2019 OI No BA SW 010 02 Granudos Touch Steuereinheit EN SCHW 8 docx Page 37 of 43 Wiring diagram buffer tank GRANUDOS 10...

Страница 38: ...conds 1 Cycle 30 seconds 30 360 seconds 30 Dosing time limit acid 0 0 100 minutes 5 Dosing time limit chlorine 0 0 100 minutes 5 Manual dosing Off Off On 2 Reserve indication chlorine Reserve message...

Страница 39: ...2 pH 6 6 7 5 pH 0 05 Delay time 5 minutes 0 60 min Activate pH Supervision like order Off On 11 Delay Booster pump Delay time 30 seconds 0 120 sec 5 12 Dosing performance Chlorine granulate rpm dosin...

Страница 40: ...on On Off On 3 Reserve indication acid Reserve message after 20l 1 500l 1 Barrel size 25l 1 500l 1 Reserve indication On Off On 4 Shock cl Dosing performance ac 10 10 100 1 Dosing performance cl 50 0...

Страница 41: ...Steuereinheit EN SCHW 8 docx Page 41 of 43 12 Dosing performance Chlorine granulate rpm dosing screw like order 12 60 RPM 1 Diameter dosing screw like order 19 26mm 1 13 Dosing performance acid Hose d...

Страница 42: ...Index 02 Date modified 11 06 2019 OI No BA SW 010 02 Granudos Touch Steuereinheit EN SCHW 8 docx Page 42 of 43 9 5 Maintenance protocol See operating instructions Part 1 Dosing technology regarding t...

Страница 43: ...earing parts are excluded from the 2 year warranty For these we assume a warranty of 1 2 year Spare parts Wear parts Device part Designation Item no Electrodes pH electrode gel with plastic shaft 1243...

Отзывы: