3.10 Replacing the inlet manifold
Requirement:
Wear clean, chemical-resistant, powder-free gloves when performing
this procedure.
Required tools and materials
• Chemical-resistant, powder-free gloves
• 1/4-inch open-end wrench
• Inlet manifold
To replace the inlet manifold:
1.
Power-off the solvent manager.
2.
Open the solvent manager's door.
3.
Disconnect the finger-tight outlet-tubing fitting from the inlet manifold outlet.
Figure 3–7: Outlet-tubing fitting on manifold
1
2
Finger-tight outlet-tubing fitting
Inlet manifold
4.
Use the 1/4-inch open-end wrench to disconnect the inlet manifold outlet-tubing fitting on
the side of the
i
2
Valve
and remove the tubing.
August 8, 2016, 715005050 Rev. C
Page 51