Figure 3–88: Pressure transducer cable in connector
1
2
Connector
Tabs
6.
Using the T27 TORX driver, loosen the 2 head bolts 1/2-turn.
Tip:
The bolts are accessible from the front of the pressure transducer.
Figure 3–89: Head bolts on accumulator pump head
1
Head bolts
Notice:
To avoid damaging the plunger, support the head from below as you
remove it, and do not tilt the head when withdrawing it.
7.
Using the T27 TORX driver, loosen and remove the 2 support-plate bolts, and then gently
withdraw the pump head and support plate from the actuator housing, making sure not to
tilt the head during the extraction.