To remove the accumulator pump head:
1.
Remove the seal-wash tubing, secured to the seal-wash housing by barbed fittings, using a
tool or by pulling on the tubing as close to the pump head as possible.
Figure 3–84: Location of seal-wash tubing
Seal-wash tubing
Seal-wash tubing
2.
Using the pliers, remove the drip wire from the head assembly.
Figure 3–85: Removing drip wire
1
2
Drip wire
Pliers
3.
Using the 1/4-inch open-end wrench, disconnect the outlet tubing from the transducer.