3-10
Maintaining the Detector
New Lamp dialog box
9. Type the serial number for the new lamp (see the label attached to the
lamp connector wire), and then click OK.
Replacing the fuses
The detector requires two 100 to 240 VAC, 50 to 60-Hz, F 3.15-A, 250-V
(fast-blow), 5 × 20 mm (IEC) fuses.
Suspect a fuse is open or otherwise defective when
•
the detector fails to power-on.
•
the fan does not operate.
To replace the fuses:
Requirement:
Replace both fuses, even when only one is open or otherwise
defective.
1. Power-off the detector and disconnect the power cord from the power
entry module.
2. Pinch the sides of the spring-loaded fuse holder, located above the power
entry module on the rear panel of the detector.
3. With minimum pressure, withdraw the spring-loaded fuse holder.
Warning:
To avoid electric shock, power-off and unplug the 2998
PDA detector before examining the fuses. For continued protection
against fire, replace fuses with those of the same type and rating
only.