C-12 Solvent Considerations
Mixed mobile phases
The table below contains approximate wavelength cutoffs for some other
solvents, buffers, detergents, and mobile phases. The solvent concentrations
represented are those most commonly used. If you want to use a different
concentration, you can determine approximate absorbance using Beer’s Law,
because absorbance is proportional to concentration.
Wavelength cutoffs for different mobile phases
Mobile Phase
UV
Cutoff
(nm)
Mobile Phase
UV
Cutoff
(nm)
Acetic acid, 1%
230
Sodium chloride, 1 M
207
Ammonium acetate,
10 mM
205
Sodium citrate, 10 mM
225
Ammonium bicarbonate,
10 mM
190
Sodium dodecyl sulfate
190
BRIJ 35, 0.1%
190
Sodium formate, 10 mM
200
CHAPS, 0.1%
215
Triethyl amine, 1%
235
Diammonium phosphate,
50 mM
205
Trifluoracetic acid, 0.1%
190
EDTA, disodium, 1 mM
190
TRIS HCl, 20 mM, pH 7.0,
pH 8.0
202, 212
HEPES, 10 mM, pH 7.6
225
Triton-X™ 100, 0.1%
240
Hydrochloric acid, 0.1%
190
Waters PIC
®
Reagent A,
1 vial/liter
200
MES, 10 mM, pH 6.0
215
Waters PIC Reagent B-6,
1 vial/liter
225
Potassium phosphate,
monobasic, 10 mM
dibasic, 10 mM
190
190
Waters PIC Reagent B-6,
low UV, 1 vial/liter
190
Sodium acetate, 10 mM
205
Waters PIC Reagent D-4,
1 vial/liter
190