Making signal connections
2-5
To connect to the electricity source:
Recommendation:
Use a line conditioner or an uninterruptible power supply
(UPS) for optimum long-term input voltage stability.
1. Place the Off/On (
) switch in the Off ( ) position.
2. Connect the female end of the power cord to the receptacle on the rear
panel of the detector.
3. Connect the male end of the power cord to a suitable wall outlet.
Important:
Do not turn the detector on at this time.
Making signal connections
See also:
Waters Ethernet Instrument Getting Started Guide
.
The following figure shows the rear panel location of the connectors used to
operate the detector with external devices.
2998 PDA Detector rear panel
B
A
NITROGEN
PURGE
RS 232
A
B
ETHERNET
!
V ~ 100 - 240
Hz 50 - 60
VA 185
F 3.15A/250V
ACN 065444751
1
2
3
4
5
6
7
8
9
10
Analog 1 (Out)
Made in USA
Serial #
EXHAUST - 15 cm (6 inch) CLEARANCE REQUIRED
!
Ground
Ground
Analog 2 (Out)
Switch 1 (Out)
Switch 2 (Out)
+
+
1
2
3
4
5
6
7
8
9
10
Inject Start (In)
Ground
Ground
Lamp Off (In)
Chart Mark (In)
Auto Zero (In)
+
+
+
+
External device
inputs and outputs