Undesirable effects
5-9
Comparison of a spectral contrast angle to its threshold angle can help
determine whether the shape difference between spectra is significant. In
general, a spectral contrast angle less than its threshold angle indicates that
shape differences are attributable to nonideal phenomena alone and that no
evidence exists for significant differences between the spectra. A spectral
contrast angle greater than its threshold angle indicates that the shape
differences arise from significant differences between the spectra. When
automating the spectral contrast comparison, the maximum absorbance of the
spectra must not exceed 1 AU.
Effects of pH on the absorbance spectrum of
p
-aminobenzoic acid
Effect of pH
pH 6.9
Wavelength (nm)
Ab
so
rbance
pH 5.1
pH 3.1