Making signal connections
2-29
Requirement:
To automatically stop the chromatographic flow to the system
in the event of an error condition or hardware failure, the Stop Flow signal
must be connected to the chromatographic pump.
Stop flow connections between the Alliance separations module and the
detector
Detector connections to an Alliance separations module
Alliance separations module
(connector B)
2424 ELS detector (connector I)
Pin 4 stop flow
+
(red)
Pin 6 stop flow
+
(red)
Pin 5 stop flow – (black)
Pin 7 stop flow – (black)
1
2
3
4
5
6
7
8
9
10
+
−
+
−
+
−
+
−
Signal Out
Signal Out
Ground
Auxiliary Out
Auxiliary Out
Stop Flow
Stop Flow
Ground
Switch Out
Switch Out
1
2
3
4
5
6
7
8
9
10
11
12
+
−
+
−
+
−
+
−
Inject Start
Inject Start
Ground
Stop Flow
Stop Flow
Hold Inject 1
Hold Inject 1
Hold Inject 2
Hold Inject 2
Ground
Chart Out
Chart Out
Alliance connector B
2424 ELS detector connector I
Red
Blacki