5-12
Diagnostic Functions and Troubleshooting
To use the gas control diagnostic
1. From the Gas & temp control menu, press 1, Gas control, to run the gas
control diagnostic. The Gas Control diagnostic screen appears.
Gas Control diagnostic screen
2. Enter the gas pressure, in psi, in the set point field.
Requirement:
Use at least 450 kPa (4.5 bar, 65 psi), to run the gas
control diagnostic.
3. To stop gas flow, enter 0 in the set point field.
Sample and reference energy diagnostic function
With the sample and reference energy diagnostic function, you view raw
sample, signal, and reference energy information, and you change the state of
the lamp and gain.
To use the sample and reference energy diagnostic function
1. Press DIAG 7, Sample & ref energy. The sample and reference energy
diagnostic screen appears.
Sample and reference energy diagnostic screen
2. Enter the gain value to view the difference between the raw sample
signal and reference energy information.
Pressure transducer
reading
Lamp energy
PMT signal level