2-6
Setting up the Detector
Power requirements
The detector, which operates over the range 100 to 240 VAC, is shipped from
the factory with two 5.00 A, 250 V fuses.
The detector’s two fuses are located above the power input receptacle on its
rear panel.
2424 ELS detector rear panel
To replace a fuse in the detector, see
Warning:
To avoid electrical shock, observe these precautions:
• Use power cord SVT-type in the United States and HAR-type (or
better) in Europe. For other countries, contact your local Waters
distributor.
• Power-off and unplug the detector before performing any
maintenance on the instrument.
• Connect all components of the HPLC system to a common ground.
Warning:
For continued protection against fire, replace fuses only with
those of the same type and rating.
TP02730
EXHAUST - 15cm (6 inch) CLEARANCE REQUIRED
RS 232
EXT 1
!
V ~ 100 - 240
Hz 50 - 60
VA 200
F 5A/250V
ACN 065444751
Signal
Auxiliary
Stop Flow
Switch
Ground
Ground
Out
Out
Out
Out
1
2
3
4
5
6
7
8
9
10
Inject Start
Lamp On
Chart Mark
Auto Zero
Ground
Ground
In
In
In
In
Made in USA
Serial #
GAS
6.9 Bar (100psi)
Maximum
EXHAUST
ETHERNET
1
2
3
4
5
6
7
8
9
10
Fuse holder
Power input
Gas inlet fitting
Inputs and
outputs
Column
heater
For service
use only
Содержание 2424
Страница 8: ...viii...
Страница 30: ...1 16 2424 ELS Detector Optics Principles...
Страница 64: ...2 34 Setting up the Detector...
Страница 110: ...3 46 Operating the Detector...
Страница 126: ...4 16 Maintaining the Detector...
Страница 182: ...A 14 Safety Advisories...
Страница 186: ...B 4 Specifications...
Страница 196: ...Index 10...