Venting the exhaust hose
2-9
Vapor trap bottle
Exhaust Venting Requirements
Requirement:
Ensure the instrument’s exhaust hose runs straight down,
toward the bench top, a minimum of 61 cm (24 inches).
•
Attach the vapor trap bottle to the end of the exhaust hose.
•
Place the vapor trap bottle’s exit hose close to an evacuation source, but
do not apply a vacuum.
Caution:
• Failure to use the vapor trap could result in too strong a vacuum,
which could adversely affect the vapor flow through the drift tube.
This could cause loss of sensitivity and excessive high-frequency
noise in the baseline.
• To avoid condensate flowing backward into the detector and thus
damaging it, run at least 61 cm (24 inches) of the instrument’s
exhaust hose vertically toward the bench top.
Barbed fittings
Bottle cap
Vapor trap bottle