Making signal connections
2-25
Note:
If the injector is an e2695 separations module running in Ethernet mode
or a 2707 autosampler, then the inject start cable should not be connected.
However, if the injector is an e2695 separations module running in IEEE
mode, then the inject start cable should be connected.
Connecting to a manual injector
If you are using a manual injector with your system, connect the signal cables
from the rear panel connector on the detector to the injector as shown in the
following table.
For information on injection trigger signals from a manual injector, see
.
Connecting to other instruments
To connect the detector to other instruments, use the two analog-out/event-in
(I/O) connectors and their corresponding pin-outs on the rear panel.
This section describes signal connections between the detector’s rear panel
and these items:
•
Waters Alliance Separations Module
•
Waters 1500-series pump
•
Waters SAT/IN™ Module
•
Waters, or other, manual injector
•
Waters autopurification system
•
Other manufacturer’s integrator or A/D interface device
Requirement:
To meet the regulatory requirements of immunity from
external electrical disturbances that can affect the performance of this
Detector connections to a manual injector
2424 ELS detector (connector II)
Manual injector
Inject Start In
+
(red)
One set of spade lug Inject Start
terminals
Inject Start In – (black)
Warning:
To avoid electrical shock, power-off the detector before making
any electrical connections.
Содержание 2424
Страница 8: ...viii...
Страница 30: ...1 16 2424 ELS Detector Optics Principles...
Страница 64: ...2 34 Setting up the Detector...
Страница 110: ...3 46 Operating the Detector...
Страница 126: ...4 16 Maintaining the Detector...
Страница 182: ...A 14 Safety Advisories...
Страница 186: ...B 4 Specifications...
Страница 196: ...Index 10...