Replacing the lamp cartridge
4-5
4. Disconnect the lamp connector from the front panel.
5. Remove the lamp cartridge assembly and replace it with a new one.
6. Reconnect the lamp connector.
7. Push the assembly back in, and tighten the two captive screws with a
Phillips screwdriver.
Warning:
To avoid electric shock, power-off and unplug the
detector before detaching the lamp power connector from the
detector.
Caution:
To avoid damaging the detector’s electronics, power-off
and unplug the detector before detaching the lamp power
connector from the detector.
Caution:
• Do not touch the new lamp with your bare fingers. Skin oils can
greatly reduce the lifetime of the lamp. If fingerprints do get on
the lamp, remove them with a lint-free tissue saturated with
ethanol.
• To avoid misaligning the lamp, do not touch the bulb height
adjustment lever.
Lamp connector