5-6
Diagnostic Functions and Troubleshooting
Tip:
To disable the Auto Gain diagnostic test, press Enter.
To disable the Auto Gain diagnostic test after it was enabled
1. Press DIAG on the detector front panel. The Diagnostic functions menu
appears.
2. Press DIAG 3, Disable Auto Gain. The Auto Gain disabled diagnostic
tests screen appears.
Auto Gain disabled diagnostic test screen
Tip:
To reenable the Auto Gain diagnostic, press Enter.
Input and output diagnostic functions
Use the input and output diagnostic functions for these purposes:
•
To view and reset the sample signal’s auto zero value
•
To fix (establish) voltage
•
To monitor two event output switches and four event input switches
To perform any of the input and output diagnostic functions, press DIAG 4,
Input & output. The Input & Output diagnostic functions menu appears.
Input & Output diagnostic functions menu
Displaying the Auto Zero offset
Using this diagnostic function, you view and reset the sample signal’s auto
zero offset value to zero.