General troubleshooting
5-17
Operational error messages
The following table contains operational error messages, along with their
descriptions, for the detector.
Reference dark
current too high
Light entering optics
bench.
Problem with PMT
or personality board.
1. Power-off and then
power-on the detector.
2. If the problem persists, call
Waters Technical Service
(see
Reference dark
current too low
Problem with PMT
or personality board.
1. Power-off and then
power-on the detector.
2. If the problem persists, call
Waters Technical Service
(see
Signal too high. Bad
PMT or light leak in
bench.
Light entering optics
bench.
1. Power-off and then
power-on the detector.
2. If the problem persists, call
Waters Technical Service
(see
System not
normalized. Run
diagnostic.
Memory was reset
and previous
normalization
information was
reset.
1. Power-off and then
power-on the detector.
2. If the problem persists, call
Waters Technical Service
(see
Operational error messages
Error message
Probable cause
Corrective action
Column heater fell
below its set
temperature
Heater failure (short
circuit).
Cable or connector
short.
Problem with
personality board.
1. Power-off and then
power-on the detector.
2. If the problem persists, call
Waters Technical Service
(see
Power-on confidence check error messages (Continued)
Error message
Possible cause
Corrective action
Содержание 2424
Страница 8: ...viii...
Страница 30: ...1 16 2424 ELS Detector Optics Principles...
Страница 64: ...2 34 Setting up the Detector...
Страница 110: ...3 46 Operating the Detector...
Страница 126: ...4 16 Maintaining the Detector...
Страница 182: ...A 14 Safety Advisories...
Страница 186: ...B 4 Specifications...
Страница 196: ...Index 10...