49
Alarms
Alarms
Alarms
Alarms
Alarm code
Alarm code
Alarm code
Alarm code
DESCRIPTION
DESCRIPTION
DESCRIPTION
DESCRIPTION
ACTION
ACTION
ACTION
ACTION
E182
E182
E182
E182
Pressure drop timeout
After the process phase the pressure took too long to drop to the atmospheric pressure. If the problem
persists call service.
E184
E184
E184
E184
Overtemperature detecded
If the problem persists call service.
E215
E215
E215
E215
Fan blocked or faulty electronic control
Call service.
E230
E230
E230
E230
Internal probe error
Switch the sterilizer OFF and ON. If the problem persists call service.
E231
E231
E231
E231
Overtemperature detecded
If the problem persists call service.
E232
E232
E232
E232----E233
E233
E233
E233----E234
E234
E234
E234
Internal probe error
Switch the sterilizer OFF and ON. If the problem persists call service.
E240
E240
E240
E240
Heating element error
Wait for the chamber to cool down. Reset the thermal overload. If the problem persists call service.
E243
E243
E243
E243
Heating element error
Reset the thermal overload. If the problem persists call service.
E310
E310
E310
E310----E320
E320
E320
E320----E33x
E33x
E33x
E33x----
E380
E380
E380
E380----E390
E390
E390
E390
Vacuum timeout
Check the door seal; clean or replace if necessary. Clean the chamber face side. Clean the chamber
filter. If the problem persists call service.
E510
E510
E510
E510
Door motor: failure after cycle completion
Switch the sterilizer OFF and ON. If the problem persists call service.
E520
E520
E520
E520
Door motor: locking timeout
If the problem persists call service.
E570
E570
E570
E570
Door motor: unable to detect the door position
Switch the sterilizer OFF and ON. If the problem persists call service.
E580
E580
E580
E580
Door motor: door locked check signal lost
If the problem persists call service.
E59x
E59x
E59x
E59x
Door motor error
Switch the sterilizer OFF and ON. If the problem persists call service.
E950
E950
E950
E950
Internal memory error
Switch the sterilizer OFF and ON. If the problem persists call service.
NOTE: Initiating a sterilization cycle is still possible.
E95x
E95x
E95x
E95x----E96x
E96x
E96x
E96x
Internal memory error
Switch the sterilizer OFF and ON. If the problem persists call service.
E990
E990
E990
E990
Manual stop
The cycle has been aborted by the user. Re-process the load.
NOTE: for any alarm not listed in this table, call technical service.