48
Alarms
Alarms
Alarms
Alarms
NOTE: for any alarm not listed in this table, call technical service.
Alarm
Alarm
Alarm
Alarm code
code
code
code
DESCRIPTION
DESCRIPTION
DESCRIPTION
DESCRIPTION
ACTION
ACTION
ACTION
ACTION
E010
E010
E010
E010
Power failure during a cycle
Load cannot be considered sterile. Repeat the cycle.
E02x
E02x
E02x
E02x
Internal voltage error
Switch the sterilizer OFF and ON. If the problem persists call service.
E041
E041
E041
E041
Cycle counter lost
Switch the sterilizer OFF and ON. If the problem persists call service.
NOTE: Initiating a sterilization cycle is still possible.
E042
E042
E042
E042
Internal clock error
Set date and time - Switch the sterilizer OFF and ON. If the problem persists call service.
NOTE: Initiating a sterilization cycle is still possible.
E060
E060
E060
E060
Internal voltage error
Disconnect optional accessories from 24VDC rear plug - switch the sterilizer OFF and ON. If the problem
persists call service.
E080
E080
E080
E080
Internal overheating
Check the dust filter and ensure that the sterilizer fan is not blocked.
E090
E090
E090
E090
Internal voltage error
Switch the sterilizer OFF and ON. If the problem persists call service.
E100
E100
E100
E100
Phase timeout
Check water level in the clean water tank. Reset the thermal overload. If the problem persists call service.
E101
E101
E101
E101
Internal probe error
Switch the sterilizer OFF and ON. If the problem persists call service.
E102
E102
E102
E102
Phase timeout
Check water level in the clean water tank. Reset the thermal overload. If the problem persists call service.
E107
E107
E107
E107
Overpressure during the pre-vacuum phase
If the problem persists call service.
E121
E121
E121
E121
Internal probe error
Switch the sterilizer OFF and ON. If the problem persists call service.
E130
E130
E130
E130
Overpressure during the sterilization phase
Clean the chamber and the chamber furniture from residuals of detergents, disinfectants and other
chemicals.
Replace the clean water if it is suspected to be contaminated with chemicals.
Ensure all the load is clean rinsed and free from any chemicals before sterilizing.
Repeat the cycle. If the problem persists call service.
E131
E131
E131
E131
Temperature fluctuation during the steril. phase
E140
E140
E140
E140
Low pressure during the sterilization phase
E150
E150
E150
E150
Low temperature during the sterilization phase
E160
E160
E160
E160
Over temperature during the sterilization phase
E163
E163
E163
E163
Overpressure detected
E180
E180
E180
E180----E181
E181
E181
E181
Internal probe error.
Switch the sterilizer OFF and ON. If the problem persists call service.