| Maintenance
54
|
Puranity TU+
VWR
Chapter 13
Step
Action
Figure
1
a.
Switch the system off (unplug main
power supply) and close the
feedater tap.
b.
Remove the ultrapure cartridge. (see
“Replacing the ultrapure cartridge”
on page 50") and replace the
RO/pretreatment cartridge. (see
“Replacing the RO/pretreatment” on
page 52)
2
a.
Click in the adapter for cleaning into
the both quick connectors for the
ultrapure cartridge.When you hear
an audible click you can be sure that
the quick-connectors have been
inserted correctly.
3
Always wear protective gloves and goggles
when handling cleaning solution.
a.
Switch the system on (connect main
power supply), open the feedwater
tap and wait until the internal 5L
tank or external tank is filled with
water and pour the contents of one
syringe of cleaning solution into the
internal 5 L tank or external tank.
Therefore open the lid of the tanks.
b.
Close the feedwater tap
NOTE!
There is no need to worry should you notice
a chlorine smell during the entire cleaning
process, as there is no risk of the limit for
chlorine gas in closed rooms being
exceeded.
a.)
Quick connectors
Adapter for
cleaning
Lid internal tank
Lid external tank
Lid
Lid
Содержание Puranity TU 12 UV+
Страница 6: ...vi Puranity TU VWR...
Страница 18: ...Extend of delivery 10 Puranity TU VWR Chapter 3...
Страница 20: ...Intended use of the device 12 Puranity TU VWR Chapter 4...
Страница 29: ...VWR Puranity TU 21 Installation Connections of the Puranity TU 3 6 system 12 11 10 9 8 7 6 1 2 3 4 5...
Страница 44: ...Flow chart 36 Puranity TU VWR Chapter 8...
Страница 46: ...How the Puranity TU system functions 38 Puranity TU VWR Chapter 9...
Страница 72: ...Equipment disposal 64 Puranity TU VWR Chapter 14...
Страница 76: ...Consumables 68 Puranity TU VWR Chapter 16...
Страница 77: ...VWR Puranity TU 69 Terminal assignment...
Страница 78: ...Terminal assignment 70 Puranity TU VWR Chapter 17...
Страница 80: ...Maintenance record 72 Puranity TU VWR Chapter 18...
Страница 84: ...76 Puranity TU VWR Index...