VWR
Puranity TU+ |
27
Installation |
Chapter 7
4
a.
Locate the RO/pretreatment
cartridge.
b.
Place the RO/pretreatment
cartridge in the front part of
the system and insert the
three quick connectors into
the RO/pretreatment
cartridge. When you hear an
audible click you can be
sure that the quick
connectors have been
inserted correctly.
NOTE!
The quick connectors are attached to
the system in such a manner so it is
to prevent installing the
RO/pretreatment cartridge
incorrectly.
5
a.
Use the R3/4“ feedwater
connecting kit, supplied with
the system, to connect the
Puranity TU+ system with
the feedwater tap. Therefore
push the R3/4“ union nut
over the outer diameter 1/4“
tube und push in the tube
into the fitting and then
connect the open end of the
outer diameter 1/4“ tube to
the feedwater connection of
the Puranity TU+ panel.
b.
After this screw in the R3/4“
union nut with the dirt
screen onto the feedwater
tap.
Step
Action
Figure
b)
RO/pretreatment cartridge
Quick connectors
a)
3/4“ Union nut
Outer diameter 1/4“ tube
Dirt screen
Fitting
Feedwater connection
Puranity TU 3+/6+
Feedwater connection
Puranity TU 12+
Содержание Puranity TU 12 UV+
Страница 6: ...vi Puranity TU VWR...
Страница 18: ...Extend of delivery 10 Puranity TU VWR Chapter 3...
Страница 20: ...Intended use of the device 12 Puranity TU VWR Chapter 4...
Страница 29: ...VWR Puranity TU 21 Installation Connections of the Puranity TU 3 6 system 12 11 10 9 8 7 6 1 2 3 4 5...
Страница 44: ...Flow chart 36 Puranity TU VWR Chapter 8...
Страница 46: ...How the Puranity TU system functions 38 Puranity TU VWR Chapter 9...
Страница 72: ...Equipment disposal 64 Puranity TU VWR Chapter 14...
Страница 76: ...Consumables 68 Puranity TU VWR Chapter 16...
Страница 77: ...VWR Puranity TU 69 Terminal assignment...
Страница 78: ...Terminal assignment 70 Puranity TU VWR Chapter 17...
Страница 80: ...Maintenance record 72 Puranity TU VWR Chapter 18...
Страница 84: ...76 Puranity TU VWR Index...