VWR
Puranity TU+ |
45
System control |
Chapter 12
Setting the limiting value for conductivity
NOTE!
The limiting value can only be set in µS/cm and is factory adjusted to a value of 0.1
µS/cm.
Setting range: 0.055 - 30 µS/cm, basic setting: 0.1 µS/cm
Step
Action
Figure
1
Press the “Menu”
button twice to enable
the limiting value for
the conductivity to be
set (red LED lights up).
2
Use the “Up/quit” and
“down” button to
increase or reduce the
value. With the “Enter”
button you switch to
the next digit. After
finishing adjustment
press “Enter” to
confirm the value.
3
Press the “Menu“
button twice to return
to the operating mode.
Display shows
Cond. Temp.
Status Sensor
Second line
Display
0.10
Limit value
or
Display shows
0.10
0.10
Display shows
Содержание Puranity TU 12 UV+
Страница 6: ...vi Puranity TU VWR...
Страница 18: ...Extend of delivery 10 Puranity TU VWR Chapter 3...
Страница 20: ...Intended use of the device 12 Puranity TU VWR Chapter 4...
Страница 29: ...VWR Puranity TU 21 Installation Connections of the Puranity TU 3 6 system 12 11 10 9 8 7 6 1 2 3 4 5...
Страница 44: ...Flow chart 36 Puranity TU VWR Chapter 8...
Страница 46: ...How the Puranity TU system functions 38 Puranity TU VWR Chapter 9...
Страница 72: ...Equipment disposal 64 Puranity TU VWR Chapter 14...
Страница 76: ...Consumables 68 Puranity TU VWR Chapter 16...
Страница 77: ...VWR Puranity TU 69 Terminal assignment...
Страница 78: ...Terminal assignment 70 Puranity TU VWR Chapter 17...
Страница 80: ...Maintenance record 72 Puranity TU VWR Chapter 18...
Страница 84: ...76 Puranity TU VWR Index...