VWR
Puranity TU+ |
55
Maintenance |
Chapter 13
4
Let the Puranity TU+ system run for 1 hour.
5
a.
Once the cleaning is completed
drain the disinfecting solution from
the internal 5L tank through the
sampling tap. Close the sampling
tap and open the feedwater tap and
run 2 tank fillings of feedwater
through the sampling tap.
b.
If you are using a Puranity TU 12+
system with external tank drain the
content of water into the drain by
open the dispensing valve on the
tank. Open the feedwater tap and
wait until the tank is filled for run
one tank filling through the sampling
tap.
c.
Switch the system off (unplug main
power supply) disconnect the
adapter for cleaning and put in the
new ultrapure cartridge. (see
“Replacing the ultrapure cartridge”
on page 50)
d.
Switch the system on (connect the
main power supply) and drain the
first 10 L through the dispensing
valve.
Step
Action
Figure
a.)
Quick connectors
Adapter for
cleaning
Sampling tap
system
b.)
Dispensing valve external
tank
c.)
Содержание Puranity TU 12 UV+
Страница 6: ...vi Puranity TU VWR...
Страница 18: ...Extend of delivery 10 Puranity TU VWR Chapter 3...
Страница 20: ...Intended use of the device 12 Puranity TU VWR Chapter 4...
Страница 29: ...VWR Puranity TU 21 Installation Connections of the Puranity TU 3 6 system 12 11 10 9 8 7 6 1 2 3 4 5...
Страница 44: ...Flow chart 36 Puranity TU VWR Chapter 8...
Страница 46: ...How the Puranity TU system functions 38 Puranity TU VWR Chapter 9...
Страница 72: ...Equipment disposal 64 Puranity TU VWR Chapter 14...
Страница 76: ...Consumables 68 Puranity TU VWR Chapter 16...
Страница 77: ...VWR Puranity TU 69 Terminal assignment...
Страница 78: ...Terminal assignment 70 Puranity TU VWR Chapter 17...
Страница 80: ...Maintenance record 72 Puranity TU VWR Chapter 18...
Страница 84: ...76 Puranity TU VWR Index...