Technical specification
NOTE!
Check at regular intervals the quality of your feedwater.
Feedwater requirements
Source
Tap water
Silt Density Index (SDI)
Bacteria should be < 0.01
Conductivity (reference tempe25 °C)
1200 μS/cm (0.0008 M
Ω
·cm)
Free chlorine
max. 0.1 ppm
Manganese content
max. 0.05 ppm
Iron content
max. 0.05 ppm
pH range
4 - 11
Temperature
+2°C - +35°C
Pressure
1 - 6 bar /14 - 87 PSI (at a pressure > 6 bar / 87 PSI a pressure
reducer must be installed upstream of the system.
Product water quality ASTM Type I (on dispensing valve)
Standard
UV
UV/UF
Conductivity
(reference tempe25 °C)
μS/cm
0.055
0.055
0.055
Resistance
(reference tempe25°C)
M
Ω
·cm
18.2
18.2
18.2
TOC value
ppb
5-10
1-5
1-5
RNase
DNase
ng/ml
pg/ul
- -
- -
- -
- -
< 0.003
< 0.4
Bacteria
CFU/ml
< 0.01
< 0.01
< 0.01
Bacterial endotoxins
EU/ml
- -
- -
< 0.001
Particles
μm/ml
< 0.2
< 0.2
< 0.2
Performance
l/min
1.0
1.0
0.6
Содержание Puranity TU 12 UV+
Страница 6: ...vi Puranity TU VWR...
Страница 18: ...Extend of delivery 10 Puranity TU VWR Chapter 3...
Страница 20: ...Intended use of the device 12 Puranity TU VWR Chapter 4...
Страница 29: ...VWR Puranity TU 21 Installation Connections of the Puranity TU 3 6 system 12 11 10 9 8 7 6 1 2 3 4 5...
Страница 44: ...Flow chart 36 Puranity TU VWR Chapter 8...
Страница 46: ...How the Puranity TU system functions 38 Puranity TU VWR Chapter 9...
Страница 72: ...Equipment disposal 64 Puranity TU VWR Chapter 14...
Страница 76: ...Consumables 68 Puranity TU VWR Chapter 16...
Страница 77: ...VWR Puranity TU 69 Terminal assignment...
Страница 78: ...Terminal assignment 70 Puranity TU VWR Chapter 17...
Страница 80: ...Maintenance record 72 Puranity TU VWR Chapter 18...
Страница 84: ...76 Puranity TU VWR Index...