VTI Instruments Corp.
50
EX10xxA Preparation for Use
d)
Click the Use the following IP address radio button.
e)
Enter the desired IP address. If using the IP from the example above,
192.168.1.1
can be
used.
f)
If not automatically completed after the IP address is entered, set the
Subnet mask
field
to
255.255.255.0
.
F
IGURE
2-11:
C
OMPLETED
NIC
S
TATIC
IP
C
ONFIGURATION
g)
Click OK to exit the network configuration properties.
T
IME
C
ONFIGURATION
The EX10xxA will initially be configured to receive its time through PTP (Precision Time Protocol).
The user can also set time using SNTP (Simple Network Time Protocol) or the time can be set
manually. The manual setting is necessary if the network environment is such that the unit cannot
reach the Internet. For more information, see
I
NPUT
C
ONNECTIONS
/
W
IRING
Thermocouple Connections
The EX10xxA employs an uncompensated (Cu-Cu) mini-thermocouple female jack as its input
connector. This connector provides a solid, reliable connection that is also easily changeable. Since
it is not thermocouple-type specific, different thermocouple types can be mixed throughout the unit
without hardware modification. The input jack is polarized and will only accept its mating connector
in one orientation. The mating connector is a standard mini-thermocouple male jack. A popular
source is the SMPW series from Omega Engineering. For specified accuracy performance, the input
connector must be of the same thermocouple type as the wire being connected.
Thermocouple wire is polarized, and it is critical to consider this polarity when connecting the
thermocouple wire to the thermocouple jack. For reference, the color designations and polarizations
of the most popular thermocouple types are listed in Table 2-2 for both ANSI (American) and IEC
(European) standards.
Содержание EX1000A
Страница 25: ...www vtiinstruments com EX10xxA Introduction 25 FIGURE 1 2 RACKMOUNT HARDWARE INSTALLATION DIAGRAM ...
Страница 27: ...www vtiinstruments com EX10xxA Introduction 27 FIGURE 1 4 ISOMETRIC AND INTERNAL CJC PLATE DIAGRAM ...
Страница 28: ...VTI Instruments Corp 28 EX10xxA Introduction EX1044 DIAGRAM ...
Страница 29: ...www vtiinstruments com EX10xxA Introduction 29 FIGURE 1 5 EX1044 TABLE TOP USAGE ...
Страница 34: ......
Страница 56: ......
Страница 74: ......
Страница 130: ......
Страница 282: ......