www.vtiinstruments.com
EX10xxA Function Calls
249
vtex10xxA_set_linear_correction
F
UNCTION
P
ROTOTYPE
ViStatus VI_FUNC vtex10xxA_set_linear_correction(ViSession
vi
, ViInt32
channels[]
, ViInt32
numChannels
,
ViReal64
gain
, ViReal64
offset
);
F
UNCTION
P
ARAMETERS
vi
= contains a session handle to the instrument. This handle is obtained by the function and remains valid until the
session is closed.
channels
= an array to receive the requested scan list. Valid input values: 0 through 47.
numChannels
= an integer output value indicating how many channels are in the scan list. Valid input values:
1 through 48.
gain
= a real input value representing the gain component,
m
, of linear EU correction, m(x-b). Valid input values:
all real numbers.
offset
= a real input value representing the offset component,
b
, of linear EU correction m(x-b). Valid input values:
all real numbers.
D
ATA
I
TEM
R
ESET
V
ALUE
For all channels:
gain
= 1.0
offset
= 0.0
D
ESCRIPTION
This function specifies m(x-b) linear correction of a voltage channel.
E
XAMPLE
Содержание EX1000A
Страница 25: ...www vtiinstruments com EX10xxA Introduction 25 FIGURE 1 2 RACKMOUNT HARDWARE INSTALLATION DIAGRAM ...
Страница 27: ...www vtiinstruments com EX10xxA Introduction 27 FIGURE 1 4 ISOMETRIC AND INTERNAL CJC PLATE DIAGRAM ...
Страница 28: ...VTI Instruments Corp 28 EX10xxA Introduction EX1044 DIAGRAM ...
Страница 29: ...www vtiinstruments com EX10xxA Introduction 29 FIGURE 1 5 EX1044 TABLE TOP USAGE ...
Страница 34: ......
Страница 56: ......
Страница 74: ......
Страница 130: ......
Страница 282: ......