VTI Instruments Corp.
42
EX10xxA Preparation for Use
F
IGURE
2-3-2:
RX10
XX
CHANNEL
SCREW
TERMINAL
PLATE
VIEW
12)
Unscrew the captive screws to inserts the thermocouple wires/ U - Lugs into it.
13)
Fix the thermocouple 'U' Lugs into screw terminals. Each channel has S,+ & - signals screws (‘S’
for Shield, ‘+’ for Positive input, ‘-‘ for Negative input). Refer the below channel layout diagram of
screw terminals.