www.vtiinstruments.com
EX10xxA Calibration
307
2)
Set the
Time Source
to “Manual”.
3)
Enable the
Set Time
checkbox.
4)
Enter the date in the
Date (MM:DD:YYYY)
fields.
5)
Enter the time in time in the
Time (HH:MM:SS)
fields using a 24-hour format.
6)
Click the
Submit
button.
6. Click submit at last
3. Enable
Set Time
2. Set to
Manual
1. Select a Time Zone
4. Set the date in MM:DD:YYYY format
5. Set the time in 24-hrs format
F
IGURE
10-5:
T
IME CONFIGURATION
P
ERFORMING
C
ALIBRATION
Once calibration equipment is connected to the EX10xxA/RX10xx and after ensuring that the time
is set properly, it can be calibrated using the SFP. The following description provides how this is
done.
1)
In the browser’s address bar, replace ‘index.cgi?’ with ‘factory.cgi?’ then press the
Enter
button. This will bring up the
Factory Mode Enable
page. Refer to the
discussion for assistance in logging into the EX10xxA/RX10xx web page.
NOTE
1) A warning may appear indicating that a factory calibration file already exists. This warning
can be ignored.
2) Calibration requires a one hour warm-up period. If the EX10xxA
/
RX10xx not has been on for
one hour, a warning may appear and the calibration process will stop at this point.
2)
When the
Enable Factory Mode
page loads, click the
Enable
button. The page will indicate
that factory mode is enabled.
F
IGURE
10-6:
EX10
XX
A
C
ALIBRATION
P
AGE
Содержание EX1000A
Страница 25: ...www vtiinstruments com EX10xxA Introduction 25 FIGURE 1 2 RACKMOUNT HARDWARE INSTALLATION DIAGRAM ...
Страница 27: ...www vtiinstruments com EX10xxA Introduction 27 FIGURE 1 4 ISOMETRIC AND INTERNAL CJC PLATE DIAGRAM ...
Страница 28: ...VTI Instruments Corp 28 EX10xxA Introduction EX1044 DIAGRAM ...
Страница 29: ...www vtiinstruments com EX10xxA Introduction 29 FIGURE 1 5 EX1044 TABLE TOP USAGE ...
Страница 34: ......
Страница 56: ......
Страница 74: ......
Страница 130: ......
Страница 282: ......