VTI Instruments Corp.
306
EX10xxA Calibration
F
IGURE
10-3:
A
CCESSING THE
EX10
XX
A
V
IA
ACE
L
OGIN INTO
SFP
EX10xxA/RX10xx calibration is performed using the SFP web interface. When accessing an action
or entry page for the first time during a session, an
Error
page will appear. To clear this error, click
on the
login
link. The
Login
page will appear. In the
Password
field, enter the configured
instrument password. The default factory-programmed password is
ex1048
. For additional
information, refer to the
NOTE
To reset the password to factory default, replace “index.cgi?” with “change_password.cgi?” in the
browser’s address bar.
F
IGURE
10-4:
L
OGIN
E
RROR
P
AGE
T
IME
C
ONFIGURATION
When calibration is performed, the time and date of calibration is recorded, making it important to
ensure that the EX10xxA/RX10xx’s time is configured properly. To determine the time, click on
the
Time Configuration
link in the command menu. If the time is displayed incorrectly, use the
following steps to configure the time from the
Time Configuration
web page.
1)
Select the correct
Time Zone
from the drop menu.
Содержание EX1000A
Страница 25: ...www vtiinstruments com EX10xxA Introduction 25 FIGURE 1 2 RACKMOUNT HARDWARE INSTALLATION DIAGRAM ...
Страница 27: ...www vtiinstruments com EX10xxA Introduction 27 FIGURE 1 4 ISOMETRIC AND INTERNAL CJC PLATE DIAGRAM ...
Страница 28: ...VTI Instruments Corp 28 EX10xxA Introduction EX1044 DIAGRAM ...
Страница 29: ...www vtiinstruments com EX10xxA Introduction 29 FIGURE 1 5 EX1044 TABLE TOP USAGE ...
Страница 34: ......
Страница 56: ......
Страница 74: ......
Страница 130: ......
Страница 282: ......