This section expands on the description of the EX10xxA’s / RX10xx’s features and explains how
to best use them.
E
NGINEERING
U
NIT
(EU)
C
ONVERSION
Each EX10xxA/RX10xx input channel can be individually configured for one of eleven different
EU conversions. The selections and their definitions are:
Conversion
Parameter
Voltage
mV
Type J
J
Type K
K
Type T
T
Type E
E
Type S
S
Type R
R
Type B
B
Type N
N
User-defined 0
User0
User-defined 1
User1
T
ABLE
3-1:
E
NGINEERING
U
NIT
C
ONVERSION
S
ETTINGS
Voltage
The instrument will return the raw voltage measured at its input
with units of volts (V). It is unaffected by the measured or input
CJC temperature for that channel.
Type J, K, T, E, S, R, B, N
The instrument will return the compensated thermocouple
temperature measured at its input with units of temperature (ºC or
ºF). The thermocouple calculations are performed using the full-
order polynomial equations and coefficients from the NIST ITS-
90 Thermocouple Database.
User-defined 0, User-defined 1
The instrument will return the compensated thermocouple
temperature measured at its input with units of temperature (ºC or
ºF). The thermocouple calculations are performed using user-
defined coefficients for the polynomial equations. More
information on this is given under
The default selection is voltage.
Содержание EX1000A
Страница 25: ...www vtiinstruments com EX10xxA Introduction 25 FIGURE 1 2 RACKMOUNT HARDWARE INSTALLATION DIAGRAM ...
Страница 27: ...www vtiinstruments com EX10xxA Introduction 27 FIGURE 1 4 ISOMETRIC AND INTERNAL CJC PLATE DIAGRAM ...
Страница 28: ...VTI Instruments Corp 28 EX10xxA Introduction EX1044 DIAGRAM ...
Страница 29: ...www vtiinstruments com EX10xxA Introduction 29 FIGURE 1 5 EX1044 TABLE TOP USAGE ...
Страница 34: ......
Страница 56: ......
Страница 74: ......
Страница 130: ......
Страница 282: ......