www.vtiinstruments.com
EX10xxA Function Calls
259
vtex10xxA_set_trigger_sourceEx
F
UNCTION
P
ROTOTYPE
ViStatus vtex10xxA_set_trigger_sourceEx(ViSession
vi
, ViInt16
vtb_masks[]
, ViInt16
dio_masks[]
, ViInt16
lan_masks[], ViBoolean
timer_enable
, ViBoolean
immediate,
ViBoolean
alarm
);
F
UNCTION
P
ARAMETERS
vi
= contains a session handle to the instrument. This handle is obtained by the function and remains valid until the
session is closed.
vtb_masks
= a return array of four 8-bit integer values representing the enabled state of trigger events from the eight
channels of the trigger bus. Within the 8-bit field, the MSB corresponds to VTB channel 7 and the LSB corresponds
to VTB channel 0. Valid input values: 0 through 255.
dio_masks
= a return array of four 8-bit integer values representing the enabled state of trigger events from the eight
channels of the digital I/O port. Within the 8-bit field, the MSB corresponds to DIO channel 7 and the LSB
corresponds to DIO channel 0. Valid input values: 0 through 255.
lan_masks
= an array of four 8-bit integer values representing the enabling of LAN events from any of the eight
LAN events. Within the 8-bit field, the MSB corresponds to LAN 7 and the LSB corresponds to LAN 0. Valid input
values: 0 (0x00h) through 255 (0xFFh).
timer
= a Boolean value indicating whether the timer is enabled as a trigger event. If this parameter is set to “1”, the
timer is enabled as a trigger. If set to “0”, timer triggering is disabled.
immediate
= a Boolean value indicating whether immediate is enabled as a trigger event. If this parameter is set to
“1”, immediate triggering is enabled. If set to “0”, immediate triggering is disabled.
alarm
= a Boolean value indicating whether the LXI alarm is used as an ARM event. If a “1” is set for this
parameter, the LXI alarm can act as an arm event. If a “0” is set, the LXI alarm is not a valid arm event.
D
ATA
I
TEM
R
ESET
V
ALUE
The trigger source is set to timer.
D
ESCRIPTION
This function sets the trigger source events. Regardless of this setting, software triggers are always enabled.
NOTE
It is necessary to explicitly call vtex10xxA_set_alarm_enable to enable timer events even when the
alarm
parameter is set to TRUE.
E
XAMPLE
Содержание EX1000A
Страница 25: ...www vtiinstruments com EX10xxA Introduction 25 FIGURE 1 2 RACKMOUNT HARDWARE INSTALLATION DIAGRAM ...
Страница 27: ...www vtiinstruments com EX10xxA Introduction 27 FIGURE 1 4 ISOMETRIC AND INTERNAL CJC PLATE DIAGRAM ...
Страница 28: ...VTI Instruments Corp 28 EX10xxA Introduction EX1044 DIAGRAM ...
Страница 29: ...www vtiinstruments com EX10xxA Introduction 29 FIGURE 1 5 EX1044 TABLE TOP USAGE ...
Страница 34: ......
Страница 56: ......
Страница 74: ......
Страница 130: ......
Страница 282: ......