Operator Manual / 6. Administration module
92
“Seasoning the tube” applies energy to warm the tube at a controlled rate.
l
Image Calibration
– Prompts a BioSans™ application to recalibrate the future scanned image data. This should
be performed by a technician only.
l
System Logs
– Launches a window showing a list of log files with the option to download them to an external
USB key. These logs are important for checking machine behavior and detecting the root causes of errors.
Figure 86 - System Logs Window
l
DIOX
– Prompts a BioSans™ application to check and verify the XR3D scanner input and output. This tool is for
technicians only.
l
Configuration Editor
– Launches the Configuration Editor tool. This is only accessible to technicians.
l
VDiag
– Launches the VDiag screen, which shows the system status and ensures if the system is ready. This is
performed by a technician or an administrator/supervisor in this module.
© VOTI 2020 / 900116-001
Содержание XR3D
Страница 8: ...Operator Manual 1 Introduction 8 Model Image XR3D 50 XR3D 60S VOTI 2020 900116 001...
Страница 9: ...Operator Manual 1 Introduction 9 Model Image XR3D 60 XR3D 6D VOTI 2020 900116 001...
Страница 10: ...Operator Manual 1 Introduction 10 Model Image XR3D 7 XR3D 7D VOTI 2020 900116 001...
Страница 11: ...Operator Manual 1 Introduction 11 Model Image XR3D 100 XR3D 100D VOTI 2020 900116 001...
Страница 12: ...Operator Manual 1 Introduction 12 Model Image XR3D 100B XR3D 100BD VOTI 2020 900116 001...
Страница 13: ...Operator Manual 1 Introduction 13 Model Image XR3D 15D XR3D 18D VOTI 2020 900116 001...
Страница 60: ...Operator Manual 5 Conducting a scan 60 Figure 51 Switching X Ray Views on Screen 2 VOTI 2020 900116 001...
Страница 96: ...Operator Manual 6 Administration module 96 Figure 91 Admin Menu Software Versions VOTI 2020 900116 001...
Страница 99: ...Operator Manual 7 TIP management 99 Figure 94 BioSans Login Menu Log in to TMS VOTI 2020 900116 001...
Страница 116: ...Accident Investigation Report Appendix...
Страница 118: ...VOTIDETECTION COM VOTI Detection 790 Begin St St Laurent QC H4M 2N5 Canada...