Operator Manual / 6. Administration module
91
l
Machine Settings
– Displays a window showing basic machine information.
Figure 85 - Machine Settings Window
l
V-View
– Prompts a BioSans™ application that maintenance staff can use to align the generator. This is
performed by a technician or an administrator/supervisor in this module.
l
Monoblock
– Prompts a BioSans™ application to season the generator. This is performed by a technician only.
Administrators and supervisors do not have access to this application.
l
X-Ray Seasoning
– Activates the application to season of the X-ray generator. If the XR3D has been powered
down for longer than 3 days, the X-ray generator tube must be seasoned to prevent damage to the tube.
© VOTI 2020 / 900116-001
Содержание XR3D
Страница 8: ...Operator Manual 1 Introduction 8 Model Image XR3D 50 XR3D 60S VOTI 2020 900116 001...
Страница 9: ...Operator Manual 1 Introduction 9 Model Image XR3D 60 XR3D 6D VOTI 2020 900116 001...
Страница 10: ...Operator Manual 1 Introduction 10 Model Image XR3D 7 XR3D 7D VOTI 2020 900116 001...
Страница 11: ...Operator Manual 1 Introduction 11 Model Image XR3D 100 XR3D 100D VOTI 2020 900116 001...
Страница 12: ...Operator Manual 1 Introduction 12 Model Image XR3D 100B XR3D 100BD VOTI 2020 900116 001...
Страница 13: ...Operator Manual 1 Introduction 13 Model Image XR3D 15D XR3D 18D VOTI 2020 900116 001...
Страница 60: ...Operator Manual 5 Conducting a scan 60 Figure 51 Switching X Ray Views on Screen 2 VOTI 2020 900116 001...
Страница 96: ...Operator Manual 6 Administration module 96 Figure 91 Admin Menu Software Versions VOTI 2020 900116 001...
Страница 99: ...Operator Manual 7 TIP management 99 Figure 94 BioSans Login Menu Log in to TMS VOTI 2020 900116 001...
Страница 116: ...Accident Investigation Report Appendix...
Страница 118: ...VOTIDETECTION COM VOTI Detection 790 Begin St St Laurent QC H4M 2N5 Canada...