Series S36 Instruction Manual
Chapter 7 Troubleshooting and Repair
800120
7-11
5.
Verify all meter configuration and troubleshooting steps previously described.
There are many possible causes of this problem, consult factory if necessary.
Symptom: Inaccurate Flow Rate Reading
1.
If the flow rate you are measuring is negative, the transducers/cables may have
been installed incorrectly. Verify the placement of the upstream and downstream
transducers and that the transducer cables are properly connected. The cable with
the red tag (upstream) must be connected to the upstream transducer and the
connector on the left of the top of the handheld unit labeled
UP
. The cable with
the blue tag (downstream) must be connected to the downstream transducer and
the connector on the right of the top of the handheld unit labeled
DN
.
2.
The first parameter to check when measuring a flow rate that is above/below the
expected flow rate of your system is the
SR Coefficient
. This can be found in the
Second Level of the Hidden Diagnostics
menu.
a.
This parameter should be set to a value of 1.0 by default. If this value is not
set to 1.0 and you have not been instructed to change it, please change it
back to a value of 1.0. This value acts as a multiplier and will either increase
or decrease the flow rate displayed on the unit.
b.
If this value is set to 1.0 and the flow rate is still above/below the expected
flow rate, continue to the next step for additional parameters to consider.
3.
The next parameter to consider is the
Speed of Sound Source
. This can be found in
the
Calibration
menu. The SonoPro® gives the user the ability to select from
three different speed of sound sources which are used in the calculation to
determine the flow rate of the liquid inside the pipe.
a.
Spacing Temp
–
This
Speed of Sound Source
uses the temperature
programmed for determining the distance (spacing) between the scribe lines
of the transducers,
Spacing Temp
, to calculate the speed of sound.
b.
Fluid Temp 1
–
This
Speed of Sound Source
uses either a manual
temperature input (
Don’t Use RTD
) or measures the temperature from an
RTD (
Use RTD
) to calculate the speed of sound.
Use RTD
is a dynamic
value and will change with the change in temperature of the fluid.
Note:
This value can only be used if the user has a model that is compatible with
external RTD inputs (i.e., VER or VERER-EM).
c.
TOF_Calc
–
This
Speed of Sound Source
is measured with the two
transducers that are installed on the pipe. This is a dynamic value and will
change with a change in the measured up and down time of flight.
4.
The next parameter to consider is the
Waveform
which can be found in the
Second
Level of the Hidden Diagnostics
menu. This parameter is used to define the
waveform of the driving circuit and can help with refining the transmit and
receive signals.
a.
Unkeyed
–
When this type of signal is selected, the driving circuit sends a
waveform in the form of a square wave with equal minimum/maximum
Содержание SonoPro S36 Series
Страница 44: ...Series S36 Instruction Manual Chapter 3 Operation 3 5 800120 Output Menu...
Страница 78: ...Series S36 Instruction Manual Chapter 7 Troubleshooting and Repair 800120 7 2 Hidden Diagnostics Menus...
Страница 92: ...Series S36 Instruction Manual Appendix A Specifications A 2 800120...
Страница 96: ...Series S36 Instruction Manual Appendix C Sound Speed and Pipe Data C 2 800120 Pipe Chart...
Страница 97: ...Series S36 Instruction Manual Appendix C Sound Speed and Pipe Data C 3 800120...
Страница 98: ...Series S36 Instruction Manual Appendix C Sound Speed and Pipe Data C 4 800120...
Страница 99: ...Series S36 Instruction Manual Appendix C Sound Speed and Pipe Data C 5 800120...
Страница 100: ...Series S36 Instruction Manual Appendix C Sound Speed and Pipe Data C 6 800120...
Страница 101: ...Series S36 Instruction Manual Appendix C Sound Speed and Pipe Data C 7 800120...
Страница 102: ...Series S36 Instruction Manual Appendix C Sound Speed and Pipe Data C 8 800120...