• The current limitation for constant current mode can be preset at output A and C with the push of a button. A short-
circuit jumper at the output is not required during setting.
• For output A and C, a voltage limitation (OVP) can be set for safety purposes. This can be set independently for
the two outputs. When this adjustment level is achieved, the respective output is switched off automatically. This
prevents the accidental destruction of connected consumers due to an excessively high output voltage.
• The device is overload and short-circuit-proof and contains a safety temperature cut-off.
• The laboratory power supply is designed in compliance with protection class 1. It is only approved for connection to
protective sockets and an alternating current of 230 V/AC commonly used in households. The earth potential socket
is directly connected to the protective earth on the mains plug.
• The mains socket to which the product is connected must be easily accessible.
• Operation under adverse ambient conditions is not permitted. Adverse conditions include: Damp or excessively
high humidity, storms or stormy conditions such as strong electrostatic fields, etc.
•
Do not use the product in environments where there are high levels of dust, flammable gases, vapours or solvents.
This may cause a fire or explosion!
• Using this product for any purpose other than those described above may damage the product and result in a short-
circuit, fire or electric shock. The product must not be modified or reassembled!
• Always observe the safety information in these instructions.