![Vitrek Xitron 2801 Скачать руководство пользователя страница 21](http://html1.mh-extra.com/html/vitrek/xitron-2801/xitron-2801_user-manual_1037509021.webp)
Functional Description
21
Current and Line Switch
Figure 4. Current and Line Switch Block Diagram
Line Switch Notes
The TRIAC and relay are controlled directly by the Channel MCU as
commanded by the DSP.
The timing of the turn on sequence when operating with AC voltages is
accomplished by the MCU delaying turning on the TRIAC (after being
commanded to do so by the DSP) by an amount commanded by the
DSP from the active edge of the “squared” signal from the comparator
in the Voltage Input circuitry. When set for DC voltage operation, the
TRIAC is turned on immediately after being commanded to do so
without reference to this signal.
Содержание Xitron 2801
Страница 1: ...USER GUIDE 2801 2802 Advanced Single and Dual Channel Power Analyzers...
Страница 2: ...2 2801 2802 Power Analyzer User Guide...
Страница 10: ...10 2801 2802 Power Analyzer User Guide...
Страница 19: ...Functional Description 19 Voltage Input Attenuator Figure 2 Voltage Input Attenuator Block Diagram...
Страница 101: ...Printing Results 101 Figure 66 Configuration Printout...
Страница 102: ...102 2801 2802 Power Analyzer User Guide Figure 67 Waveform Channel 1 Real time...
Страница 103: ...Printing Results 103 Figure 68 Waveform Channe1 Distortion...
Страница 104: ...104 2801 2802 Power Analyzer User Guide Figure 69 History Chart VOLTS ACDC Trends View...
Страница 105: ...Printing Results 105 Figure 70 History Chart AMPS PERCENT THD Extents View...
Страница 106: ...106 2801 2802 Power Analyzer User Guide Figure 71 Harmonics Listing page 1...
Страница 107: ...Printing Results 107 Figure 72 Harmonics Listing page 2...
Страница 145: ...Sending Commands and Interrogatives 145...
Страница 147: ...Appendix A Physical Specifications 147...
Страница 153: ...Appendix B Measurement Specifications 153...