Appendix C-Input & Target Specifications 157
Transducer (15V) Input Specifications
Input Burden
50k
Ω
|| 10pF, shield grounded via 50
Ω
Ranges 15V
(35Vpk)
5V
(12.5Vpk)
1.5V
(3.5Vpk)
0.5V
(1.25Vpk)
0.15V
(0.35Vpk)
Max. Input
70Vrms, 200Vpk continuous (any range)
Scaling
Measurements may be scaled by a user entered factor
of between 0.001 and 10000. The result is scaled into
the units of Amps.
Max. External
Offsets of up to 6V may be compensated for any
DC Offset
range, compensation is performed at the input prior to
range
selection.
Compensation
When enabled, the input may be polarity reversed
and/or adjusted to compensate for the current caused
by the measured voltage across the Voltage Input
Burden.
RMS Accuracy
As shown in the charts for AC, AC+DC and DC only
parameters,
filtered
(20kHz) and unfiltered.
Rectified Accuracy
1.5x RMS accuracy
Form Factor Accuracy 0.5x RMS accuracy, as a percentage of 1.0
Peak Accuracy
Filtered (20kHz) : RMS ac 2mV + 0.01% of
peak
range.
Unfiltered : RMS ac 10mV + 0.02% of peak
range.
Crest Factor Accuracy Filtered (20kHz) : (2mV + 0.01% of peak range)
/RMS
value
Unfiltered : (10mV + 0.02% of peak range)/RMS
value.
Harmonics Accuracy Absolute: (RMS accuracy) * (1 + (h/100))
Relative: (RMS accuracy) * (h/100)
Floor : <0.02% of range (any single harmonic),
<0.035% of range (all harmonics).
Содержание Xitron 2801
Страница 1: ...USER GUIDE 2801 2802 Advanced Single and Dual Channel Power Analyzers...
Страница 2: ...2 2801 2802 Power Analyzer User Guide...
Страница 10: ...10 2801 2802 Power Analyzer User Guide...
Страница 19: ...Functional Description 19 Voltage Input Attenuator Figure 2 Voltage Input Attenuator Block Diagram...
Страница 101: ...Printing Results 101 Figure 66 Configuration Printout...
Страница 102: ...102 2801 2802 Power Analyzer User Guide Figure 67 Waveform Channel 1 Real time...
Страница 103: ...Printing Results 103 Figure 68 Waveform Channe1 Distortion...
Страница 104: ...104 2801 2802 Power Analyzer User Guide Figure 69 History Chart VOLTS ACDC Trends View...
Страница 105: ...Printing Results 105 Figure 70 History Chart AMPS PERCENT THD Extents View...
Страница 106: ...106 2801 2802 Power Analyzer User Guide Figure 71 Harmonics Listing page 1...
Страница 107: ...Printing Results 107 Figure 72 Harmonics Listing page 2...
Страница 145: ...Sending Commands and Interrogatives 145...
Страница 147: ...Appendix A Physical Specifications 147...
Страница 153: ...Appendix B Measurement Specifications 153...