95x Series Operating Manual - May 17, 2022
Page
88
of
155
0ohm at 10A : 3.5uohm
0ohm at 40A : 1.25uohm
TEST TIMING
Ramp
0.00 to 9999sec
<(± 1% ± 0.1sec ± 1 cycle) accuracy
Dwell
0.02 to 9999sec or user terminated
<(± 0.05% ± 20ms ± 1 cycle) accuracy
Fast Discharge
<(20ms + 1 cycle)
GROUND LEAKAGE TESTING (DCI AND ACI)
These types are intended for testing that a product does not exceed ground leakage requirements, usually of
safety standards.
Both types are similar, the difference being that the DCI type tests for DC ground leakage current while the ACI
type tests for AC ground leakage current.
ACTIONS WHILE RUNNING
During the test step the 95x measures the ground leakage current and checks that the result is within the user set
range. The 95x also measures the voltage on the
HV
terminal relative to ground and makes the result of this
measurement available to the user, if not required then this can be ignored.
Optionally, the user can also detect arc currents and fail the DUT if they are found over the user set limit.
The user can set for there to be a delay at the beginning of the test step before the limits are enforced on the
ground leakage current measurement.
CONFIGURING
An example DCI menu is as follows
–
Seq 1 Step 1 DCI
TIME: 25.0sec
Delay: 0.05sec
Lim: 0n-35.00mA
ARC DETECT:4us 10mA
ON FAIL: ABORT SEQ
An example ACI menu is as follows
–
Seq 1 Step 1 ACI
TIME: 25.0sec
Delay: 0.05sec
Lim: 0n-35.00mA
ARC DETECT:4us 10mA
ON FAIL: ABORT SEQ
•
TIME
. Allows the user to program the total test time.
•
Delay
. Allows the user to set a delay time before which the limits are not enforced.
•
Lim
. Allows the user to set a minimum and maximum limit for the ground leakage current beyond which
the DUT will be failed.