95x Series Operating Manual - May 17, 2022
Page
68
of
155
100V
10K-60G
Ω
: ±0.7%±(R/2G
Ω
)%
10K-25G
Ω
:
±0.7%±(R/800M
Ω
)%
10K-15G
Ω
:
±0.7%±(R/500M
Ω
)%
250V
10K-150G
Ω
: ±0.6%±(R/5G
Ω
)%
10K-40G
Ω
: ±0.6%±(R/1.3G
Ω
)%
10K-25G
Ω
:
±0.6%±(R/750M
Ω
)%
500V
17K-300G
Ω
:
±0.6%±(R/10G
Ω
)%
17K-60G
Ω
:
±0.6%±(R/1.75G
Ω
)%
17K-30G
Ω
:
±0.6%±(R/900M
Ω
)%
1000V
34K-600G
Ω
:
±0.6%±(R/20G
Ω
)%
34K-75G
Ω
: ±0.6%±(R/2G
Ω
)%
34K-35G
Ω
: ±0.6%±(R/1G
Ω
)%
2500V
84K-1.5T
Ω
: ±0.6%±(R/50G
Ω
)%
84K-80G
Ω
: ±0.6%±(R/2.5G
Ω
)%
84K-40G
Ω
: ±0.6%±(R/1G
Ω
)%
5000V
167K-3T
Ω
:
±0.6%±(R/100G
Ω
)%
167K-90G
Ω
:
±0.6%±(R/2.5G
Ω
)%
167K-40G
Ω
: ±0.6%±(R/1G
Ω
)%
10000V
1.34M-6T
Ω
:
±0.6%±(R/200G
Ω
)%
1.34M-90G
Ω
:
±0.6%±(R/2.5G
Ω
)%
1.34M-40G
Ω
:
±0.6%±(R/1G
Ω
)%
957i
100V
10K-60G
Ω
: ±1.2%±(R/2G
Ω
)%
10K-20G
Ω
:
±1.2%±(R/500M
Ω
)%
10K-10G
Ω
:
±1.2%±(R/300M
Ω
)%
250V
25K-150G
Ω
: ±1.1%±(R/5G
Ω
)%
25K-30G
Ω
: ±1.1%±(R/1G
Ω
)%
25K-15G
Ω
:
±1.1%±(R/500M
Ω
)%
500V
50K-300G
Ω
:
±1.1%±(R/10G
Ω
)%
50K-40G
Ω
: ±1.1%±(R/1.3G
Ω
)%
50K-20G
Ω
:
±1.1%±(R/650M
Ω
)%
1000V
100K-600G
Ω
:
±1.1%±(R/20G
Ω
)%
100K-50G
Ω
:
±1.1%±(R/1.5G
Ω
)%
100K-25G
Ω
:
±1.1%±(R/700M
Ω
)%
2500V
250K-1.5T
Ω
:
±1.1%±(R/50G
Ω
)%
250K-60G
Ω
:
±1.1%±(R/1.75G
Ω
)%
250K-25G
Ω
:
±1.1%±(R/800M
Ω
)%
5000V
500K-3T
Ω
:
±1.1%±(R/100G
Ω
)%
500K-60G
Ω
:
±1.1%±(R/1.8G
Ω
)%
500K-25G
Ω
:
±1.1%±(R/800M
Ω
)%
10000V
1M-6T
Ω
: ±1.1%±(R/200G
Ω
)%
1M-60G
Ω
: ±1.1%±(R/1.8G
Ω
)%
1M-25G
Ω
:
±1.1%±(R/800M
Ω
)%
15000V
3M-10T
Ω
: ±1.6%±(R/300G
Ω
)%
3M-60G
Ω
: ±1.6%±(R/1.8G
Ω
)%
3M-25G
Ω
:
±1.6%±(R/800M
Ω
)%
CURRENT MEASUREMENTS (DUT GROUNDED, OPTION HSS ONLY)
Measurements are performed in the
HV
terminal of the 95x.
Breakdown Current
1uA to 50mApk (DC-50kHz bandwidth)
<(± 1% ± 1uA) accuracy
<30usec detection time
<3msec response time (typically <500usec)
Leakage Current
0.00uA to 50mA
7.25msec measurement period (100ms for dwell time >2 sec)
<(± 1% ± 50nA ± (10nA per KV)) accuracy
Capacitive Loading Effects :
Add X * (± 0.2nA ± (2.5nA per KV))
X = (C/0.01uF) below 0.01uF, or X = √(C/1uF) above 1uF, or otherwise X = 1.0
Cable Compensation
<(± 1% of error)
CURRENT MEASUREMENTS (DUT GROUNDED, OPTION HSS-2 ONLY)
Measurements are performed in the
HV
terminal of the 95x.